Code: IU.01.09.0045
Version/Revision: 2.000
Page: 33 - 55
Drying cabinet
Model: EA-34-03
Monthly
• Retighten the electrical contact screws.
• Check wear of the door trim.
Annually
• Recalibrate the temperature sensor.
• Carry out a general inspection of the entire electrical and mechanical system.
• Check system grounding conditions.
Cleaning external finishes:
• Clean with a soft cloth or sponge (non-abrasive), wet in hot or warm water and neutral soap;
• If in doubt about the cleaning materials that will be used, consult your supplier regarding their application
to stainless steel and acrylic.
Cleaning the bowl surface:
If the bowl has encrusted material, spray the surface with neutral detergent:
• Wait a few minutes and scrub the bowl with a non-metallic brush (a brush with nylon bristles is ideal).
Remove the product with rinse and/or a damp cloth;
• Be careful not to leave lint on the surface.
When carrying out these routines, always use appropriate PPE.
Make sure the equipment is turned off.
Never use an abrasive sponge, steel wool or metal instruments that could damage the surface.
Cleaning the external finishes and the interior of the chamber must be done using a soft fabric
and solutions that are not aggressive to stainless steel, painted carbon steel, ABS, glass and
acrylic.
Some surfaces may have a high temperature. Before starting cleaning, wait for the temperature
to drop to safe values (less than 40°C).
9.1. Wear Items
Below are the components that suffer natural wear and tear due to the use of the equipment, which is why these
items are not covered by the product's Limited Warranty:
• Component trims;
• Valve repairs;
• Filter and filter elements;
• Dot matrix printer ribbon;
• Pump seals;
• Bearings, pulleys, journals and couplings;
• Relays and Contactors;
• Lubricating oils.
9.2. Other important recommendations
The recommendations presented below must be followed, as they are important for the safety of operators and
equipment, as well as for its correct functioning, as some are provided for in specific standards:
• Instruments and controls must be calibrated and checked for good operating conditions annually;
• It is recommended that the revalidation process is carried out in all its phases and operators are trained.
Such procedures, in addition to ensuring the longevity of the product, guarantee those responsible for