the connection tube of the EASYSPRAY device is supplied without fittings. The
appropriate fitting must be mounted to the end of the supply line by technical service
personnel.
Prior to first time use, the protective caps have to be removed from the gas connections.
The recommended pressurized gas source is compressed air or nitrogen.
CAUTION: Pressurized gas supply must be between 3.5 and 7 bar.
4.2 PREPARING FOR OPERATION
The device should be visually inspected for any mechanical damage that could impair
the function at regular intervals.
If the low-battery indicator light is illuminated, change battery immediately.
If the device is no longer safe in function and/or operation, discontinue use and return
it to Baxter in accordance with Section 7 (RETURN OF DEVICES).
4.3 OPERATION
The clamp on the back side of the EASYSPRAY device allows the device to be
mounted onto an IV-pole or a bed-rail. Firmly set screws to assure stable connection
to either the IV-pole or bed-rail.
Connect the supply line to pressurized gas outlet in the OR (compressed air or
nitrogen).
CAUTION: Pressurized gas supply must be between 3.5 and 7 bar.
NOTE: Follow instructions for use for surgical sealant and spray device preparation.
Connect the clear sterile filter of the spray set pressure line to the male luer connector
on the EASYSPRAY device. Connect the blue filter of the spray set sensing line to
the blue marked female luer connector on the EASYSPRAY device.
Turn the switch on the front side of the EASYSPRAY device to the ON-position (I).
Adjust the pressure on the EASYSPRAY device according to the specifications in the
spray set instructions for use by turning the regulator knob in the appropriate
direction.
To activate the flow of gas, cover the sensing line outlet (the hole on the spray set clip)
as described in the spray set instructions for use.
The quantity of product applied is controlled by the product application system but
not by the gas released.
When spraying of sealant has been completed or the EASYSPRAY device is no
longer in use, turn the switch on the front side to the OFF-position (O).
CAUTION: For the recommended spraying distance, gas pressure and application
details read the spray set instructions for use carefully.
6
G4505_EasySpray_IFU_US_3OK 24.05.2005 15:06 Uhr Seite 6

Other manuals for baxter EASYSPRAY

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter EASYSPRAY and is the answer not in the manual?

baxter EASYSPRAY Specifications

General IconGeneral
Brandbaxter
ModelEASYSPRAY
CategoryMedical Equipment
LanguageEnglish

Related product manuals