Overpressure Shut-Down:
Inspection of the overpressure shut-down mechanism:
1. Connect the EASYSPRAY device to the pressurized gas supply.
2. Switch the On/Off switch into the On position (I).
3. Use the pressure control knob to set a pressure of approx. 2.8 bar.
4. Attach a calibration manometer (calibrator) with measuring range of at least 4 bar
to the pressure line connector.
5. Dismount the pressure regulator knob with the aid of an Allen wrench (2.5 mm).
6. Remove the retaining washer.
7. Mount the pressure control knob.
8. Close/Cover the sensing line connector to start gas flow through the EASYSPRAY device
and set the output pressure to 2.8 bar.
9. Check the function of the EASYSPRAY device by starting it repeatedly, i.e., covering
and uncovering the sensing line connector.
10.Using the pressure regulator increase the output pressure by 0.1 bar.
11.Repeat points 9 and 10 until gas flow through the device is interrupted (shut-down
pressure has been reached).
12.The shut-down pressure must be between 3.1 bar and 3.7 bar (3.4 +/- 0.3 bar).
13. Switch the On/Off switch to Off (O).
14. Dismount the pressure control knob.
15. Reinstall the retaining washer.
16. Mount the pressure control knob.
Battery Inspection:
1. Open the battery compartment with the aid of a coin or screwdriver and remove the
battery from drawer.
2. Check the battery compartment and the drawer for contaminants.
3. Check the contact springs for corrosion.
4. Test the battery voltage (using a battery tester if necessary) or dispose of the battery
in accordance with the local regulations.
5. Re-insert the tested or new battery into the drawer and test the device for function.
6. TERMS OF GUARANTEE
The manufacturer guarantees that all defects in material and fabrication occurring
within 2 years after the date of sale are repaired free of charge.
This guarantee can only be claimed under the following conditions:
• Inform Baxter immediately of malfunction. Please provide the serial number of the device
when reporting.
• Observe all instructions regarding storage or return of the device.
• Present a legible copy of the invoice for the device in question, which clearly shows
the date of sale.
• Describe as exactly as possible the suspected defects or malfunctions by the user.