2. WARNINGS AND PRECAUTIONS
CAUTION: Pressurized gas supply must be between 3.5 and 7 bar.
CAUTION: DO NOT connect to an oxygen source. Failure to follow these instructions
can lead to explosion, which could result in serious injury or death.
Prior to operation make sure that the pressure range of the device is adjusted to the
range specified in the appropriate spray set instructions for use (See 4.1 FIRST-TIME
USE).
CAUTION: Any application of pressurized gas may be associated with a potential risk
of air embolism, tissue rupture or gas entrapment with compression, which may be
life threatening. Be sure to take appropriate measures to address these risks by
observing the recommended minimum spraying distance and maximum pressure provided
in the appropriate spray set instructions for use.
CAUTION: Spraying into enclosed body cavities requires appropriate safety measures
to make sure that the above mentioned risks will be avoided.
ATTENTION:
DO NOT throw or drop the device!
Do not use the device in the case of visible damage caused by improper handling. Return
damaged devices to Baxter for repair.
3. STORAGE
Storage and Transport Conditions: Temperature -10º to 45ºC, 60% relative air
humidity.
Protect from heat, dust, humidity, frost and direct sunlight.
If the device is not used on a regular basis, observe the following:
Prior to each re-use, perform a functional test (See 4.1 FIRST-TIME USE).
Connect the device to the pressurized gas supply and vent at the maximum pressure
by repeatedly closing the sensing line luer connector.
4. OPERATION
4.1 FIRST-TIME USE
After removing the device from the package, it is recommended to keep the carton for
potential maintenance or repair. Check the device visually for any mechanical damage
that could impair operation.
Open the battery compartment with the aid of a coin or screwdriver and insert a
standard 9 Volt battery. NOTE: Do not use rechargeable batteries.
Since operating rooms are equipped with various types of fittings for pressurized gas,
5
G4505_EasySpray_IFU_US_3OK 24.05.2005 15:06 Uhr Seite 5

Other manuals for baxter EASYSPRAY

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter EASYSPRAY and is the answer not in the manual?

baxter EASYSPRAY Specifications

General IconGeneral
Brandbaxter
ModelEASYSPRAY
CategoryMedical Equipment
LanguageEnglish

Related product manuals