BD Phoenix™ M50 System Quick Reference Guide
W201 Missing Accession Number Operation was attempted and value in Accession # field was invalid or blank. Type in/scan correct
accession number before attempting operation again.
W202 Unknown Accession Number Attempt was made to find data or print report for accession number that is not in the BD Phoenix™
database. Verify that correct accession number was entered. If error recurs, record may need to be
logged in if it is new.
W203 Missing Isolate Number Attempt was made to save record when accession number is present without isolate number. If an
accession number is entered, an isolation number must also be entered.
W204 Missing Media Type Attempt was made to save a Yeast ID panel during Panel Login and no Media is specified.
W205 Panel Not Found Attempt was made to find a Sequence and Accession combination that is not in the BD Phoenix™
database. The Sequence and Accession combination does not exist in the database.
W300 Missing Test Strain for a QC
Panel
Attempt was made to save a QC panel without organism ID (Test Strain). An organism (Test Strain) must
be selected to save a QC panel.
W301 Missing Tech ID for a QC Panel Attempt was made to save a QC panel without Tech ID. A tech ID must be entered to save a QC panel.
W302 Invalid Tech ID for a QC Panel Attempt was made to save a QC panel with invalid Tech ID. Enter up to 3 alphanumeric characters
excluding: * ? [ ] ! # | for Tech ID.
W303 Invalid Lot Number Attempt was made to save record/print report for invalid panel lot number, ID or AST broth lot number, or
indicator lot number. Enter correct lot number as shown on the carton.
W304 Missing Lot Number Attempt was made to save record/print report for a QC panel without panel lot number, or when
expiration date was entered for an optional lot number field with no corresponding lot number. A panel
lost number must be entered to save a QC panel. If an expiration date is entered for optional lot number
field, a corresponding lot number must also be entered.
W305 ID or AST Must Be Enabled Attempt was made to save a panel/QC panel, and both ID and AST checkboxes are unchecked. At least
one option must be enabled for panel record to be saved and testing to occur.
W400 Door Already Open A request was made to perform a task that requires the door to be opened when the door is already
open. Select OK.
Code Possible Cause
Corrective Action
(If the recommended actions do not correct the situation, refer to the
BD Phoenix™ M50 User’s Manual or contact BD)
18
Error Codes W201 – W400