Checking a program
1. Switch on the unit. The current temperature of
the furnace is displayed after approx.
3 seconds.
2. Select program:
Press as often until LED # lights up.
Press until desired program number
appears.
3. Check values:
Press to check the values of
heating-up speed (°C/min),
holding temperature (°C) and
holding time (min.).
4. To check the next holding level, go to the
field using the buttons and call up
the next holding level using the
buttons.
5. Check the settings at this holding level as
described in point 3.
6. The current furnace temperature is displayed
again approx. 10 seconds after the last button
was pressed. The check is now finished.
NOTE
The current furnace temperature is displayed
again approx. 10 seconds after the last button
was pressed. This also happens when hesitat-
ing too long before pressing the next key.
Always check all holding levels! All values in the
unused holding levels of a program must be set
to "0“.
Changing values: page 38.