Replacement of a defective thermocouple
Time and/or program are no
longer stored
Controller defective Replace controller.
LED lights up, but furnace
does not get warm
Heater defective Check heater for continuity.
If necessary, replace heating
mould (→ page 49).
Controller defective Replace controller.
Furnace does not get warm when
"Set Finished time" has been set
Long power failure Check in-house power network
for time switching.
"Reducing or increasing temper-
ature in running program" does
not function
The "Start/Stop" button was
pressed.
* Create new program with desired
final temperature. Start program.
Calculated and actual Finished
time deviate significantly from
each other
Voltage fluctuations in the power
network
* Carry out correction program
(→ page 42).
Jamb brick defective Replace jamb brick (→ page 50)
1)
If an error message occurs (Er XX), an acoustic signal sounds for 3 minutes.
Malfunction Cause Elimination
page 47