34
Displays / Error Messages
Function Message Cause
Rs, IPSC
RI, IPSC
Rs, IPSC
RI, IPSC
Rs, IIPSC
RI, IPSC
RCD
Rs, IPSC
RI, IPSC
RCD
Rs, Ik; RI, IPSC;
RCD
PE - Error
Rs, IPSC
RI, IPSC
RCD
RCD
RCD
RCD
RCD has not tripped.
RCD
RCD
RCD
RCD
RCD
Loop impedance is higher than 1999 Ohm
Contact voltage has exceeded preset limit during the socket test.
Measurement therefore not carried out, until instrument has been
allowed to cool down.
TELARIS 0100 is overheated, therefore no
further measurements can be carried out.
Socket incorrectly connected, or RCD has tripped.
Insert plug after turning it by180°C or change L and N of test leads.
Trip time exceeds 500ms
Trip time exceeds 1999 ms or RCD has not tripped
Trip time exceeds 50 ms or RCD has not tripped
Trip time exceeds 300 ms or RCD has not tripped
Selective RCD has tripped after 19 ms.
RCD error, trip time is less than 60 ms
Measurement of trip time Ia
(only at nominal fault current I∆N = 5 x 30 mA)
(only at nominal fault current I∆N = 0.5 x 30 mA)
Test of selective RCD´sTest of selective RCD´s
Measurement of trip current, Display of trip current
For correct measuremet touch contact electrode
For correct measurement touch contact electrode
Contact-
electrode has
touched