SECTION SEVEN : INITIAL SETTING
[1] FILM SPEED
As factory installation, following three kinds of lm speed are registered to be selected by
Film Speed Selection Switch :
a = Film speed No. F.08 (equivalent to ISO speed group "D", or Kodak Ultra-Speed lm)
b = Film speed No. F.03 (equivalent to ISO speed group "F/E", or Kodak InSight lm)
c = Film speed No. F.02 (equivalent to ISO speed group "F")
Including these three, Model 096 can provide 16 different types of lm speed and any three of them
can be registered for easy selection.
If the doctor uses different speed of lm, or prefers darker (or lighter) radiograph, substitute speed
can be registered as follows:
1. While depressing technical switch
, turn on the main power switch
. Film type lamp "a" is lit,
and F.09 is displayed in exposure time display window
.
Then, release technical switch.
2. By depressing
switch (or switch), increase (decrease) lm speed number until desired
number is displayed. [Refer to Exposure Time Table on next page.]
3. Depress technical switch
, an electronic chime sounds and the selected lm speed number is
registered at lm type "a".
4. Turn off the main power switch.
5. If different lm speeds are to be registered at "b" and "c", depress "F"switch
after step 1 above
to light the appropriate lm type lamp, and repeat steps 2 & 3.