[2] PRIORITY OF SELECTION
As factory installation, following selection lamps light when the main power switch is turned on:
Patient Size : Medium
Cone : Regular
Film Speed : "a"
If necessary, this priority can be changed as follows;
[For example, at pedodontistry, patient size of "small" should be preferentially selected.]
1. While depressing technical switch
16
, turn on the main power switch.
2. Select the patient size "small" by depressing patient size selection "P"switch
10
.
3. Depress technical switch
16
, an electronic chime sounds and the patient size of "small" will be
stored as primary selection.
4. Priority of selection for cone type and lm speed can be changed by same procedures.
[3] ELECTRONIC CHIME ON/OFF
As factory installation, electronic chime sounds when each switch is depressed. If preferred, this
sound can be eliminated.
1. While depressing tooth selection switches T1 & T2 together, turn on the main power switch.
2. "bu.1" will be displayed in exposure time display window
13
.
3. By depressing either or (
3
or
4
), display changes to "bu.0".
4. Then depress technical switch
16
, and turn off the main switch.
NOTE :
Exposure Warning Buzzer and alarm sound of error code can not be eliminated.
NOTE :
At step 3 of [1] FILM SPEED (page 26) and [2] PRIORITY OF SELECTION (page 28),
occasionally " 0.00" will be displayed in Exposure Time Display Window when
technical switch
16
is depressed. In that case :
1. Turn OFF the Main Power Switch.
2. Turn ON the Main Power Switch Again.
3.
Conrm that Patient Size, Cone Type and Film Speed are correctly selected.
4. Press Film Speed Selection Switch and conrm F.xx displayed in Exposure Time Display
Window is correct.