Recommended safety distance between portable and mobile RF telecommunication devices and the BEMER PRO SETS
The BEMER-SET PRO is designed for operation in an electromagnetic environment in which the RF disturbances are controlled. The customer
or the user of the BEMER-SETS PRO can help to prevent electromagnetic disturbances by keeping a minimum distance between portable
and mobile RF telecommunication devices (transmitters) and the BEMER-SET PRO – depending on the power output of the communication
device, as indicated below.
OUTPUT POWER
OF THE TRANSMITTER W
SAFETY DISTANCE, DEPENDENT ON THE TRANSMITTED FREQUENCY m (IN METERS)
150 KHZ TO 80 MHZ 80 MHZ TO 800 MHZ 800 MHZ TO 2.7 GHZ
d=1.2 P d=1.2 P d=2.3 P
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters whose maximum output power rating is not listed in the table above, the distance can be determined by calculating
the equation on each column, where P is the maximum output power rating of the transmitter in watts (W) according to the trans-
mitter manufacturer.
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any part of the BEMER Professional device, including cables specified by the manu-
facturer. Otherwise, degradation of the performance of this equipment could result.
1960-2002-BEDIENANL-B.BOX-USA-FDA-2020-US-en-v13-Akku Rev20_4.indd 451960-2002-BEDIENANL-B.BOX-USA-FDA-2020-US-en-v13-Akku Rev20_4.indd 45 04.11.2020 08:47:3104.11.2020 08:47:31