Electromagnetic immunity
The BEMER SET CLASSIC / PRO is designed for operation in the electromagnetic environment listed below.
The customer or user of the BEMER SETS CLASSIC / PRO should ensure that the device is used in such an environment.
EMISSION TESTS IEC 60601 TEST LEVEL COMPLIANCE LEVEL ELECTROMAGNETIC ENVIRONMENT GUIDELINES
Electrostatic discharge according to IEC 61000-4-2 ± 6 kV Contact discharge
± 8 kV Air discharge
± 6 kV Contact discharge
± 8 kV Air discharge
Fast transient electrical interferences according to IEC 61000-4-4 ± 2 kV for power supply lines ± 2 kV for power supply lines
The quality of the power supply should correspond to that of a typical commercial or hospital environment.
Surge voltage / surges according to IEC 61000-4-5 ± 1 kV Voltage outer conductor – outer conductor
± 2 kV Voltage outerconductor – earth
± 1 kV Voltage outer conductor – outer conductor
± 2 kV Voltage outerconductor – earth
The quality of the power supply should correspond to that of a typical commercial or hospital environment.
Voltage dips, interruptions and variations of the power supply according to
IEC 61000-4-11
< 5% U
T
for 1/2 cycles (> 95 % dip in U
T
)
< 40 % U
T
for 10 cycles (60 % dip in U
T
)
<70 % U
T
for 25 cycles (30 % dip in U
T
)
< 5 % U
T
5 s (> 95 % dip in U
T
)
< 5 % U
T
for 1/2 cycles (> 95 % dip in U
T
)
< 40 % U
T
for 5 cycles (60 % dip in U
T
)
<70 % U
T
for 25 cycles (30 % dip in U
T
)
< 5 % U
T
5 s (> 95 % dip in U
T
)
The quality of the power supply should correspond to that of a typical commercial or hospital environment.
Power frequency (50 / 60 HZ) magnetic field according to IEC 61000-4-8 3 A/m 3 A/m
Note: U
T
is the alternating power supply voltage prior to application of the test level