REV: 13/05/2022 · ©2022 Benewake (Beijing) Co., Ltd. | en.benewake.com | All rights reserved
BP-UM-30 A01
13
Figure 5-2 BW_TFDS software interface and display
① Please turn off any anti-virus software before uncompressing the PC
software. Otherwise, the software maybe detected as virus and deleted
by anti-virus. The software is only runnable under Windows.
② If no data is available in area “④TIME LINE CHART”, please check the
wire connection and sequence. When TF02-Pro is successfully powered
on, there will be a red indicator light inside transmitting lens viewing
from the front.
③ The value of distance output Dist may vary with the output unit, which
is cm by default. If the unit of distance is changed to the unit-mm with
specific command, and the PC software will be unable to identify it, and
so the unit of “④TIME LINE CHART” will still be cm. For example, the
actual TF02-Pro measurement is 1m, the distance value of TF02-Pro is
1000 in mm, the value read by the PC software also is 1000, but the unit
will not change and still display cm.