CHROMOPHARE® D 650plus, D 530plus, D 500plus, D 650, D 530, D 500 Service Manual (E)
15.POWER REQUIREMENTS
15.1 List of fuse specifications
Fuse values to be installed on-site [A]
220 – 240 V110 – 120 V
LightL
F1
[A]
Battery
F2/F3
[A]
L
F1
[A]
Battery
F2/F3
[A]
D 500plus
D 530plus
210410
D 650plus4166,316
D 500
D 530
210410
D 6504166,316
All fuse slow blow
15.2 List of terminal voltages
The terminal voltages specified in table below must be provided on the flanged tube of the CHROMOPHAR
light or the Power Box
Ceiling tubePower Box
LightA.C.
[V]
D.C.
[V]
A.C.
[V]
D.C.
[V]
D 500plus
D 530plus
21,5 (*24 + 1,5/-0,321,5 (*24 + 1,5/-0,3
D 650plus21,5 (*24 + 1,5/-0,321,5 (*24 + 1,5/-0,3
D 500
D 530
21,5 (*24 + 1,5/-0,321,5 (*24 + 1,5/-0,3
D 65021,5 (*24 + 1,5/-0,321,5 (*24 + 1,5/-0,3
(* filtering board next in line