Questions and Answers

  • J
    Jessica CarrollSep 12, 2025
    What to do if the voltage of the control channel deviates from the admissible set value in Berthold Measuring Instruments?
    • L
      Lisa WilliamsSep 12, 2025
      If the voltage of the control channel deviates from the admissible set value in your Berthold Measuring Instruments, restart the device. If the event occurs frequently, contact BERTHOLD.
  • M
    Matthew GordonSep 9, 2025
    What to do if the voltage of the measuring channel deviates from the admissible set value in Berthold Measuring Instruments?
    • K
      Kathy EvansSep 9, 2025
      If the voltage of the measuring channel deviates from the admissible set value in your Berthold Measuring Instruments, restart the device. If the event occurs frequently, contact BERTHOLD.
  • K
    Karen HoffmanSep 8, 2025
    How to fix an error in the real-time clock of Berthold Duo Measuring Instruments?
    • M
      Mr. Richard RobbinsSep 8, 2025
      If there is an error in the real-time clock of your Berthold Measuring Instruments, reset the date and time, and restart the device. The event must be confirmed manually.
  • T
    Tina RussellSep 7, 2025
    What to do if the internal detector temperature is close to the limit of Berthold Duo?
    • A
      Angela RiveraSep 7, 2025
      If the internal detector temperature of your Berthold Measuring Instruments is too hot: Install/check water cooling system. If the detector is too cold: Install/check heating.
  • A
    Amanda SanchezAug 23, 2025
    What to do if there is a hardware error with Berthold Duo Measuring Instruments?
    • B
      Brooke AdkinsAug 23, 2025
      If a hardware error occurs with your Berthold Measuring Instruments, it indicates a defect. Contact BERTHOLD for assistance. The hardware may need to be checked and replaced.
  • J
    jesusewingSep 5, 2025
    What happens if proper function of Berthold Duo can no longer be guaranteed?
    • D
      Daniel WilliamsSep 5, 2025
      If the proper function of the Berthold Measuring Instruments can no longer be guaranteed, it is recommended to have the device checked by Berthold Technologies, even if it still seems to work properly.
  • M
    Mr. Jeffrey VasquezSep 4, 2025
    What to do if the current high voltage deviates from the average high voltage by more than 20% in Berthold Duo Measuring Instruments?
    • L
      Latoya RodriguezSep 4, 2025
      If the current high voltage deviates from the average high voltage by more than 20% in your Berthold Measuring Instruments, check the detector electronics and photomultiplier. If necessary, contact BERTHOLD.
  • B
    Brett WhiteSep 3, 2025
    What causes the watchdog to trigger a restart of Berthold Measuring Instruments?
    • S
      Stephen BerrySep 3, 2025
      If the watchdog has triggered a restart of the Berthold Measuring Instruments, and the event occurs frequently, contact BERTHOLD. Verify whether massive electromagnetic disturbances have triggered the event.
  • R
    Ruben LittleSep 2, 2025
    What to do if there is an error in the high-voltage supply of the photomultiplier in Berthold Duo?
    • F
      froblesSep 2, 2025
      If an error in the high-voltage supply of the photomultiplier was detected in your Berthold Measuring Instruments, check the detector electronics and photomultiplier. If necessary, contact BERTHOLD.
  • A
    Amber KingAug 31, 2025
    What to do if the average HV is more than 20% lower than the Default-HV in Berthold Duo Measuring Instruments?
    • W
      williamsjamesAug 31, 2025
      If the average HV is more than 20% lower than the Default-HV in your Berthold Measuring Instruments, check the Default HV for plausibility and change it if necessary. Also, check the detector electronics and photomultiplier. If necessary, contact BERTHOLD.

Summary

Safety

Dangers and Safety Measures

Essential instructions for safe operation and handling of the product.

Proper Use

Defines correct usage of the detectors and outlines improper usage scenarios.

Qualification of the Personnel

Details required qualifications for personnel handling the device.

Operators Obligations

Lists the operator's responsibilities for personnel training and product monitoring.

System Description

Measuring Principle

Explains the radiometric measurement principle and system components.

Installation

Safety Instructions

Critical safety warnings and precautions for installation procedures.

Installation of Mounting Clamps

Guidance on how to install mounting clamps for detectors.

Electric Installation

Initial Start-up

Step-by-step guide for the initial electrical setup and connection of the detector.

Operation

PC Software

System requirements and connection details for using the PC software.

Operating the PC Software

Explanation of the PC software's structural design and control options.

The PC Software Menu

Schematic structure of the detector menu within the PC software.

Reset

Procedures for restarting the detector or resetting to factory settings.

Troubleshooting

System Events

List of possible device event messages and troubleshooting options.

Maintenance and Repair

Safety Instructions

Safety precautions for maintenance and repair work, especially in hazardous areas.

Software Update

Procedure for updating the device software using the Detector Service Modem.

Replacing the Entire Detector

Steps for removing and installing a complete detector unit.

Replacing the Electronics Module

Instructions for replacing the detector's electronics module.