LaserSpeed 8000/9000 I/O Module Instruction Handbook
Setup
Part No. 93342 / Drawing No. 0921-01516 Page 57 of 84 Revision E (Dec 2013)
~T Gauge Type
~T<CR> Read Value
Response Format:
GaugeType
Type ls9k
~T<X><CR> Set Value
Factory Default: ls9k
Where <X> is the type of gauge connected to the I/O Module. This setting should
be set to “ls9k”.
Example: To configure the I/O Module to communicate with LS8000/LS9000
gauges, send the command ~Tls9k<CR>.
~Y Master/Slave Connection Status
~Y<CR> Read Value
Response Format:
Master:<X1> Slave:<X2>
Where <M> indicates if a Master is connected, and <S> indicates if a Slave is
connected.
This command returns the connection state of the rear panel Sync In and Sync
Out connectors. The I/O Module detects whether or not a Master and/or Slave is
connected. This information is used to calculate Elongation, and to control the
function of the Length/Elongation parallel port if Auto-Elongation Mode
(~I:Elongation setting).
<X1> and <X2> are each either Connected or None.
See the Elongation/Synchronization section for details about Elongation.