EasyManua.ls Logo

Binder ED 56 - User Manual

Default Icon
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Issue 02/2022 Art. No. 7001-0292
Operating Manual
BD (E3.1)
|
Incubators
Avantgarde.Line with natural convection
BF (E3.1)
|
Incubators
Avantgarde.Line with forced convection
ED (E3.1)
|
Drying and heating ovens
Avantgarde.Line with natural convection
FD (E3.1)
|
Drying and heating ovens
Avantgarde.Line with forced convection
FED (E3.1)
|
Drying and heating ovens
Avantgarde.Line with forced convection
and enhanced timer functions
with microprocessor temperature controller
Model Model version Art. No. Model Model version Art. No.
BD 56
BD056-230V 9010/ 9110-0323
ED 260
ED260-230V 9010/ 9110-0339
BD056UL-120V 9010/ 9110-0324 ED260UL-240V 9010/ 9110-0340
BD 115
BD115-230V 9010/ 9110-0325 ED 720 ED720-400V 9010/ 9110-0341
BD115UL-120V 9010/ 9110-0326
FD 56
FD056-230V 9010/ 9110-0303
BD 260
BD260-230V 9010/ 9110-0329 FD056UL-120V 9010/ 9110-0304
BD260UL-120V 9010/ 9110-0330
FD 115
FD115-230V 9010/ 9110-0305
BD 720
BD720-230V 9010/ 9110-0331 FD115UL-120V 9010/ 9110-0306
BD720UL-240V 9010/ 9110-0332
FD 260
FD260-230V 9010/ 9110-0309
BF 56
BF056-230V 9010/ 9110-0313 FD260UL-240V 9010/ 9110-0310
BF056UL-120V 9010/ 9110-0314 FD 720 FD720-400V 9010/ 9110-0311
BF 115
BF115-230V 9010/ 9110-0315
FED 56
FED056-230V 9010/ 9110-0295
BF115UL-120V 9010/ 9110-0316 FED056UL-120V 9010/ 9110-0296
BF 260
BF260-230V 9010/ 9110-0319
FED 115
FED115-230V 9010/ 9110-0293
BF260UL-120V 9010/ 9110-0320 FED115UL-120V 9010/ 9110-0294
BF 720
BF720-230V 9010/ 9110-0321
FED 260
FED260-230V 9010/ 9110-0299
BF720UL-240V 9010/ 9110-0322 FED260UL-240V 9010/ 9110-0300
ED 56
ED056-230V 9010/ 9110-0333
FED 720
FED720-400V 9010/ 9110-0301
ED056UL-120V 9010/ 9110-0334 FED720UL-208V 9010/ 9110-0302
ED 115
ED115-230V 9010/ 9110-0335
ED115UL-120V 9010/ 9110-0336
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: customerservice@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Other manuals for Binder ED 56

Questions and Answers

Summary

Safety

Intended Use and Risk Management

Proper applications, foreseeable misuse, and residual risks of the chamber.

Installation Procedures

Electrical Connection

Detailed information on electrical connections, including cables and fuses.

Start Up Procedures

Controller Operation and Settings

Set-Point and Air Flap Adjustment

How to set temperature/fan speed and adjust the air flap position.

Password Management

Procedure for setting or changing user and controller access passwords.

Overtemperature Protection

Protective Devices and Controllers

Description of Class 1 protective device and Class 2/3.1 safety controllers.

Alarm Handling and Function Checks

Procedures for handling alarms and checking safety controller functionality.

Timer Functions

Ramp Function

Ramp Function Basics and Setup

Explanation of temperature ramps and step-by-step guide to setting them up.

Data Recording via USB

Starting and Terminating Data Recording

Steps to initiate and stop data recording using a USB stick.

Network Configuration

Network Settings (MAC, DHCP, IP, Subnet, Gateway)

Procedures for configuring network parameters like MAC address, DHCP, IP, Subnet, and Gateway.

Cleaning and Decontamination

Cleaning Procedures and Agents

Procedures and recommended agents for cleaning chamber surfaces and components.

Decontamination Methods

Methods for decontaminating the chamber after exposure to hazardous substances.

Maintenance, Service, and Troubleshooting

Personnel and Maintenance Schedule

Requirements for personnel and recommended schedule for regular maintenance.

Troubleshooting and Returning for Service

Guide to resolving common faults and procedures for returning the chamber.

Disposal Procedures

Summary

Safety

Intended Use and Risk Management

Proper applications, foreseeable misuse, and residual risks of the chamber.

Installation Procedures

Electrical Connection

Detailed information on electrical connections, including cables and fuses.

Start Up Procedures

Controller Operation and Settings

Set-Point and Air Flap Adjustment

How to set temperature/fan speed and adjust the air flap position.

Password Management

Procedure for setting or changing user and controller access passwords.

Overtemperature Protection

Protective Devices and Controllers

Description of Class 1 protective device and Class 2/3.1 safety controllers.

Alarm Handling and Function Checks

Procedures for handling alarms and checking safety controller functionality.

Timer Functions

Ramp Function

Ramp Function Basics and Setup

Explanation of temperature ramps and step-by-step guide to setting them up.

Data Recording via USB

Starting and Terminating Data Recording

Steps to initiate and stop data recording using a USB stick.

Network Configuration

Network Settings (MAC, DHCP, IP, Subnet, Gateway)

Procedures for configuring network parameters like MAC address, DHCP, IP, Subnet, and Gateway.

Cleaning and Decontamination

Cleaning Procedures and Agents

Procedures and recommended agents for cleaning chamber surfaces and components.

Decontamination Methods

Methods for decontaminating the chamber after exposure to hazardous substances.

Maintenance, Service, and Troubleshooting

Personnel and Maintenance Schedule

Requirements for personnel and recommended schedule for regular maintenance.

Troubleshooting and Returning for Service

Guide to resolving common faults and procedures for returning the chamber.

Disposal Procedures

Binder ED 56 Specifications

General IconGeneral
BrandBinder
ModelED 56
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals