EasyManua.ls Logo

Binder FED 720 - User Manual

Default Icon
117 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Issue 02/2022 Art. No. 7001-0292
Operating Manual
BD (E3.1)
|
Incubators
Avantgarde.Line with natural convection
BF (E3.1)
|
Incubators
Avantgarde.Line with forced convection
ED (E3.1)
|
Drying and heating ovens
Avantgarde.Line with natural convection
FD (E3.1)
|
Drying and heating ovens
Avantgarde.Line with forced convection
FED (E3.1)
|
Drying and heating ovens
Avantgarde.Line with forced convection
and enhanced timer functions
with microprocessor temperature controller
Model Model version Art. No. Model Model version Art. No.
BD 56
BD056-230V 9010/ 9110-0323
ED 260
ED260-230V 9010/ 9110-0339
BD056UL-120V 9010/ 9110-0324 ED260UL-240V 9010/ 9110-0340
BD 115
BD115-230V 9010/ 9110-0325 ED 720 ED720-400V 9010/ 9110-0341
BD115UL-120V 9010/ 9110-0326
FD 56
FD056-230V 9010/ 9110-0303
BD 260
BD260-230V 9010/ 9110-0329 FD056UL-120V 9010/ 9110-0304
BD260UL-120V 9010/ 9110-0330
FD 115
FD115-230V 9010/ 9110-0305
BD 720
BD720-230V 9010/ 9110-0331 FD115UL-120V 9010/ 9110-0306
BD720UL-240V 9010/ 9110-0332
FD 260
FD260-230V 9010/ 9110-0309
BF 56
BF056-230V 9010/ 9110-0313 FD260UL-240V 9010/ 9110-0310
BF056UL-120V 9010/ 9110-0314 FD 720 FD720-400V 9010/ 9110-0311
BF 115
BF115-230V 9010/ 9110-0315
FED 56
FED056-230V 9010/ 9110-0295
BF115UL-120V 9010/ 9110-0316 FED056UL-120V 9010/ 9110-0296
BF 260
BF260-230V 9010/ 9110-0319
FED 115
FED115-230V 9010/ 9110-0293
BF260UL-120V 9010/ 9110-0320 FED115UL-120V 9010/ 9110-0294
BF 720
BF720-230V 9010/ 9110-0321
FED 260
FED260-230V 9010/ 9110-0299
BF720UL-240V 9010/ 9110-0322 FED260UL-240V 9010/ 9110-0300
ED 56
ED056-230V 9010/ 9110-0333
FED 720
FED720-400V 9010/ 9110-0301
ED056UL-120V 9010/ 9110-0334 FED720UL-208V 9010/ 9110-0302
ED 115
ED115-230V 9010/ 9110-0335
ED115UL-120V 9010/ 9110-0336
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: customerservice@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Other manuals for Binder FED 720

Questions and Answers

  • M
    Mr. Luke WatsonSep 12, 2025
    What to do if my Binder Laboratory Equipment chamber doesn’t heat up and the red heating control light is off?
    • J
      James ScottSep 12, 2025
      If your Binder Laboratory Equipment chamber isn’t heating up, the red heating control light is off, but the controller display is working, the timer may have run off. Program the timer or switch to continuous operation.
  • D
    dscottSep 12, 2025
    Why is my Binder FED 720 Laboratory Equipment chamber heating permanently?
    • W
      William CookSep 12, 2025
      If your Binder Laboratory Equipment chamber is heating permanently and the set-point isn't being held, the controller might be defective, so it's recommended to contact BINDER service. The Pt 100 sensor might also be defective. Calibrate and adjust the controller.
  • E
    Erin OconnorSep 12, 2025
    What to do if my Binder FED 720 Laboratory Equipment set-point temperature is not reached?
    • M
      Mary NelsonSep 12, 2025
      If the set-point temperature isn't reached in your Binder Laboratory Equipment after the specified time, ensure the chamber door is completely closed. If that doesn't solve the issue, the door gasket may be defective and need replacement. Also, calibrate and adjust the controller. Verify that the power supply matches the voltage requirement of 115V or 230V.
  • R
    Rachel PhelpsSep 9, 2025
    Why won't my Binder Laboratory Equipment chamber heat up, but the red heating control light is on?
    • C
      crystal94Sep 10, 2025
      If your Binder Laboratory Equipment chamber isn't heating up, and the red heating control light is on, the heating element may be defective, so it's recommended to contact BINDER service. Also, a semiconductor relay might be defective.
  • V
    vclarkSep 8, 2025
    What to do if my Binder FED 720 temperature is too high and the red alarm light is on?
    • C
      Cindy HoustonSep 9, 2025
      If the temperature inside your Binder Laboratory Equipment chamber is too high and the red alarm light is on, check the settings of the temperature set-point and of the safety device class 3.1.
  • R
    Rachel HaleSep 7, 2025
    What to do if my Binder Laboratory Equipment chamber has no function?
    • K
      Kayla SmithSep 8, 2025
      If your Binder Laboratory Equipment chamber isn't working, first check the connection to the power supply. Also, inspect the chamber fuse and replace it if needed. If the fuse blows again, it is recommended to contact BINDER service.
  • S
    Sandra SmithAug 23, 2025
    What to do if Binder FED 720 Laboratory Equipment chamber has no function?
    • A
      angelica97Aug 23, 2025
      First, check the connection to the power supply. If that's not the issue, the overtemperature protective device class 1 might have turned off the chamber. Disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, further assistance is needed.
  • D
    Deborah ParkerSep 5, 2025
    Why are there deviations from the indicated heating-up times in my Binder Oven?
    • C
      Christina LeonardSep 6, 2025
      Deviations from the indicated heating-up times in your Binder Oven can occur if the chamber is fully loaded. Reduce the load or allow longer heating-up times.
  • M
    Mr. Paul Gutierrez Jr.Sep 4, 2025
    Why does the fan in my Binder Oven not turn or turns too slowly?
    • M
      Michael MooreSep 5, 2025
      If the fan in your Binder Oven doesn’t turn or turns too slowly, first check if the fan speed is set too low and set the fan speed to 100%. If the problem persists, contact Binder service.
  • M
    Megan FullerSep 4, 2025
    Why is my Binder Oven chamber without function, with only the green stand-by LED lit?
    • M
      Morgan MillerSep 4, 2025
      If your Binder Oven chamber has no function and only the green “stand-by” LED is lit, the chamber is likely in stand-by mode. Press the ON/OFF button until the display lights up.

Summary

1. Safety

1.1 Personnel Qualification

Requirements for personnel operating the chamber, including training and authorization.

1.4 Structure of the Safety Instructions

Explanation of safety symbols, signal words, and warning panels.

1.7 General Safety Instructions on Installing and Operating

Broad safety advice for chamber installation, operation, and ventilation.

1.8 Intended Use

Defines the proper applications and limitations of the chamber.

2. Chamber Description

2.2 Triangular Instrument Panel

Details the control panel, its display, and functional buttons.

3. Completeness of Delivery, Transportation, Storage, and Installation

3.1 Unpacking and Checking Delivery

Steps for verifying the received chamber and its accessories.

3.2 Guidelines for Safe Lifting and Transportation

Safety precautions and procedures for moving the chamber.

3.4 Location of Installation and Ambient Conditions

Requirements for the chamber's installation site and environmental conditions.

4. Installation

4.2 Electrical Connection

Detailed steps for connecting the chamber to the power supply.

6. Overview and General Settings on the R4 Controller

6.1 Controller Overview

Explanation of controller buttons, display, and status symbols.

6.6 Set-point Entry for Temperature and Fan Speed

How to set target temperature and fan speed parameters.

6.8 Changing Passwords for User Level and General Controller Functions

Procedure to set or change access passwords for controller functions.

7. Overtemperature Protection

7.1 Overtemperature Protective Device (Class 1)

Description of the basic overtemperature safety device.

7.2 Safety Controller

Details on the adjustable electronic safety controller.

7.3 Setting the Safety Controller Set-point

How to configure the limit temperature for the safety controller.

7.4 Alarm Message and Proceeding in Case of an Alarm

Actions to take when an alarm is triggered by safety devices.

8. Timer Functions

8.1 Selecting the Timer Function

Overview of the three available timer modes: Delayed Off, Temp. Dependent Delayed Off, Delayed On.

8.2 Timer Function “Delayed Off”

Instructions for setting the delayed-off timer.

8.2.1 Entry and Activation of Timer Run-time and Fan Setting

How to input timer duration and fan settings for Delayed Off.

8.3 Timer Function “Temperature Dependent Delayed Off” (BF, FED)

Steps for setting a temperature-dependent timer.

8.3.1 Entry and Activation of Timer Run-time, Fan Setting and Set-point Entry

Inputting parameters for the temperature-dependent timer.

8.4 Timer Function “Delayed On” (BF, FED)

Steps for setting the delayed-on timer.

8.4.1 Entry and Activation of Timer Run-time and Fan Setting

Inputting parameters for the delayed-on timer.

9. Ramp Function

9.2 Setting and Displaying the Ramp Function

How to configure and view ramp settings, gradient, and target set-point.

10. Data Recording via USB Interface

10.1 Starting Data Recording

Steps to initiate data logging to a USB stick.

10.2 Terminating Data Recording

Steps to stop data logging and retrieve recorded data.

13. Cleaning and Decontamination

13.1 Cleaning

Procedures for cleaning chamber surfaces and components.

13.2 Decontamination; Chemical Disinfection

Methods for disinfecting the chamber after contamination.

14. Maintenance and Service, Troubleshooting, Repair, Testing

14.3 Simple Troubleshooting

Basic steps to resolve common operational issues.

Binder FED 720 Specifications

General IconGeneral
BrandBinder
ModelFED 720
CategoryLaboratory Equipment
LanguageEnglish