Manual for VMP300-based instruments
2, 3, 4 or 5 terminal leads (+ ground)
Gain-Bandwidth compensation
9 programmable stability factors
Highest unity gain bandwidth
Adjustable from ± 10 V down to ± 30 mV (1 mV
resolution)
± 10 V, 300 µV resolution.
For example: For an Erange of [0;20]V, the resolution is
333.33 µV. For [0;19.66]V, the resolution is 300 µV. For
[0;13.1]V, the resolution is 200 µV. For [0; 6.55]V, the
resolution is 100 µV. For [0; 3.27]V, the resolution is 50 µV.
For [0; 1.31]V, the resolution is 20 µV. For [0; 0.65]V, the
resolution is 10 µV. For [0; 0.32]V, the resolution is 5 µV.
For [0; 0.13]V, the resolution is 2 µV. For [0; 0.06]V, the
resolution is 1 µV.
< ± 1 mV ± 0.03% of setting
± 1 A, ± 100 mA, ± 10 mA, ± 1 mA, ± 100 µA,
± 10 µA, ± 1 µA (7 ranges + Autorange)
± 100 nA, 10 nA with gain
< ± 0.1 % of range ± 0.03 % of setting
± 10 V, ± 5 V, ± 2.5 V, ± 250 mV, ± 25 mV
± 10 V, 300 µV resolution
< ± 1 mV, ± 0.03 % of reading
Filtering low-pass 4 poles Sallen-
Key filters
± 1 A, ± 100 mA, ± 10 mA, ± 1 mA, ± 100 µA,
± 10 µA, ± 1 µA
< ± 0.1 % of range ± 0.03 % of reading
Filtering low-pass 4 poles Sallen-
Key filters