11. TECHNICAL SPECIFICATIONS & PERFORMANCES
11.1 ELECTRICAL RATINGS
CAUTION
Make sure you connect the Mallya USB cable to a power source compliant with the
IEC 60601-1.
USB CHARACTERISTICS
Input voltage
5V DC 2A
Min Current
500mA
Ingress Protection
IP 22
BATTERY CHARACTERISTICS
Battery
Li-ion rechargeable :
Li-ion 3,7V-130mAh
Model Number
GEB401730 / FT401235P
Operating time with fully charged battery
2 weeks
Charging time for fully charged battery
Full charge in 2.5 hours
Table 17 - Mallya electrical information
You can connect the supplied USB cable to the USB port of your PC or to an AC adapter
suitable for your region, with the following characteristics:
AC ADAPTOR (NOT PROVIDED)
External Module with automatic adaptation of the voltage: no selection is required.
Input voltage range
90-264 V
Frequencies range
47-63 Hz
Category
II (double insulation)
Output voltage
5V DC
Output current
500mA 2A
Table 18Battery charging information
11.2 MATERIALS
MALLYA PART
PEN MODEL
MATERIALS
Mallya base All
PC/ABS (cover)
Copolyester (body)
Mallya button
FLEXPEN
®
NOVO NORDISK
SOLOSTAR
®
SANOFI
Polypropylène and
POM
KWIKPEN
®
ELI LILLY
TPE
Table 19 - Materials
11.3 ACCURACY
MEASUREMENT ACCURACY
99%
Table 20Accuracy on test bench
CAUTION
External factors may affect the dose measurement accuracy.
Factors Affect the measurement
Does not affect the
measurement
Ambient Temperature
Magnetic disturbance
Table 21 - External factors and accuracy of the displayed value
To display the accurate doses, please follow the recommended conditions of use.

Table of Contents