Handheld Pulse Oximeter
7 Precautions for Use
Precautions for Electric Safety
8
PalmCare plus
Standard operational conditions are as follows.
•Temperature : 10℃ ~ 45℃ (50 ~ 113℉)
•Altitude : 70 ~ 106kpa
•Humidity : 30 ~ 85%
Standard storage and transportation conditions are as follows.
Temperature : -10℃ ~ 50℃ (14 ~ 122℉)
Altitude : 70 ~ 106kpa
Himidity : 20 ~ 95%
The product may not work well when it is kept or operated beyond the temperature and
humidity specified.
If you want to turn off the PalmCare plus while you’re operating, please use the power switch
where is the front of the product.
When you keep or operate the product in large temperature difference, product
malfunctioning can be happened because of humidity of battery pack.
Please operate product after keep it room temperature for three hours.
Leakage, heating, ignition or rupture of the battery may cause fire or injury.
Do not use batteries beyond those specified in this manual.
Do not short-circuit the battery and do not heat, disassemble, or dispose of it in a fire.
•Do not mix using the used battery with new batteries, different kind of batteries, other
company batteries
When the product is not used for a long time, separate the battery from the product.
To ensure the lifecycle and safety of battery, please discharge the battery and then charge
perfectly again every six months. Also time-worn battery exceed 24 months should be
replaced.
When the product has been moved, make sure to turn it off and arrange the accessories in
order. Injured cords or cables may cause fire or electric shock.
Prior to operation of the equipment, make sure to check following itmes.
①Whether power supply is appropriate. (AC100V~AC220V)
②Whether connections (power line or selected equipment) are properly made to the equipment.
③Whether Probes are connected before the systems is turned on.
Precautions for Electric Safety
The equipment should be placed far from generator, X-ray equipment, broadcasting
equipment, or transmitting wires, so as to prevent the electrical noises form being generated
during the operation, when these devices are near the equipment, it can produce inaccurate
measurements, For Palmcare plus , both independent circuit and stable grounding are
essentially required. In the event that same power source is shared with other electronic
equipment, it can also produce inaccurate output
Palmcare plus is classified as follows;
It is a Class I, type-BF against electric shock.
Equipments are not suitable for use in presence of a flammable anaesthetic mixture and a
solvent.
NOTE
CAUTION CAUTION
CAUTION