Handheld Pulse Oximeter
9 Description of Palmcare plus
How to Set PalmCare plus
10
PalmCare plus
Description of Palmcare plus
Characteristics and Principles of PalmCare plus
PalmCare plus is pulse oximeter for measuring oxygen saturation of arterial blood and plus rate of
patient. Data and detected wave are displayed on the front LCD.
There are two types of probe to the Plamcare plus. One is reusable probe and the other is disposable
probe to apply neonate or pediatric patient as well as single use to adult.
There are two principles that pulse oximeter is based on.
Oxyhemoglobin and Deoxyhemoglobin differ in spectral absorption, the relative absorption of red
and infrared light by bold reflects hemoglobin oxygen saturation. And the changed volume of
arterial blood in tissue could be measured during the pulsatile cycle.
A pulse oximeter measures the oxygen saturation of arterial hemoglobin by passing red and infrared
light emitting diodes(LED) in the oximeter sensor serve as light source (ultraviolet rays 660nm and
infrared light 960nm); a photodiode serves as the photo detector.
Pulse rate and arterial blood oxygen saturation are pleased on a LCD display in the real time. Users
can determine the alarm settings and volume as needed via cont key panel on the front. PalmCare
plus gives the operator the ability to measure patient basic vital biometric data.
Options Accessories
①Disposable probe
②Extension cord 2.4m(8ft)
③Cradle
④USB(mini B TYPE) serial cable
Composition of PalmCare plus(Accessories)
Main 1 set
Adapter(1EA)
DC 5V 2A
SpO2 Finger Probe (1EA)
Operation Manual
(1EA)
Standard Accessories
1.1 Precautions for Settings
Pay attention when setting Palmcare plus for the following points
- Use Palmcare plus under conditions of 10℃ 〜 40℃ of circumstantial temperature and 30% 〜
85% humidity.
- Check the connection status of the power cord.
- Do not connect multiple cords to the power supply.
- Place the main body on a flat area.
- If noise occurs, ground the equipment.
- Do not use any electric cords generating connecting noise.
- Take care since the product can be broken by mechanical shock.
- Get rid of dust or inflammable material near the product when setting it up.
1.2 Checkpoints before Use
Check the following points before measuring the patient’s condition.
- Make sure there is no mechanical risk.
- Check the lead, power plug, and accessories connected on the outside.
◎
Precautions Before Use
①Do not use product near dynamo, X-ray equipment, broadcast equipment.
②Do not twist the cable.
③Attach each Lead properly.
④Keep the patient motionless with comfortable position while measuring
1.3 Maintenance and Cleaning After Use
A. Keep the product cleans by cleaning it using a soft cloth at least once a month. Do not use thinner,
ethylene, oxidant or lacquer since these may damage the product.
B. Keep the accessories clean of dust or foreign materials. For cleaning, use a soft cloth and clinical
alcohol.
C. Do not soak the accessories in liquid or detergents. Keep the product and accessories out of any
liquid.
D. For the protection of the environment, please contact the agency you purchased the product from
when you wish to discard the Palmcare plus.
E. Our product can be used semipermanently. Please contact our Customer Service dept. or any
distributor and then replace the parts, when wasting parts such as battery.
1. How to Set PalmCare plus
F. Regular preventive maintenance should be carried out annually.
You are responsible for any requirements specific to your country.