BIOQUELL PROTEQ USER MANUAL -EN
TD092-O&M-001
Page 28 of 58
Uncheck
Check to allow the injection of H
2
2
during dwell phase. Timed cycle
Uncheck
Check allows the aeration phase to
finish after the aeration time has
been completed and will continue, if
required, until concentration is below
1ppm. If unchecked the aeration will
run for the set time only.
Rate*
4 to 12 (with
Flexi Adaptor)
2
2
injection rate during gassing
phase.
2
2
vaporised in gassing phase. (timed
cycle only, gassing parametric
Room volume to be decontaminated
(gassing parametric enabled)
4 to 12 (with
Flexi Adaptor)
2
2
injection rate during dwell
phase.
Length of time in dwell phase
At the end of gassing dwell the
system checks that the peak PPM
reached thus far is greater than the
alert level. If not a warning alarm
message will be displayed and
Delay time before the volt free cycle
phase indicator switches from
gassing to aeration. Delay time must
be shorter than aeration time
Length of aeration phase if Aeration
Parametric is unchecked, and
minimum aeration time if checked.
Number of additional wireless
aeration units which are required.
* These values must be set with care to ensure hydrogen peroxide does not
condense in the vaporiser. Seek advice from Bioquell.