User Supplied Parts
This section will describe any parts that you will need to provide, in or-
der to properly operate the equipment.
1. Qty. 1 regulator, per compressed gas source, either a one stage regulator or a two
staged regulator. Either way you will need two gauges on the regulator, one to monitor the
amount of gas in the tank and one to monitor the gas coming out of the tank. (2500 PSIG
input 0-60 PSIG output, recommended). PSIG stands for Pounds per Square Inch Gauge
as opposed to PSIA which stands for Pounds per Square Inch Absolute.
2. Qty. 1 compressed gas source (tank) of nitrogen. (If using for low oxygen levels).
3. Qty. 1 compressed gas source (tank) of oxygen. (If using for high oxygen levels).
5
Proox model 360