J
Jason ReyesAug 21, 2025
How to fix error 2311 to 2386 on Biotek ELx808IU Laboratory Equipment?
- Cchelsea63Aug 21, 2025
This error indicates a defective or old lamp. Replace the lamp to resolve the issue.
How to fix error 2311 to 2386 on Biotek ELx808IU Laboratory Equipment?
This error indicates a defective or old lamp. Replace the lamp to resolve the issue.
What does error 0A00 mean on Biotek ELx808IU?
Error 0A00, 0A01 to 0A05 indicates that the lamp has failed since power-up. Check that the lamp is lit and replace it if necessary.
What to do if I get error 0900 on my Biotek ELx808IU?
Error 0900 can occur if the read chamber door has opened or closed relative to when the unit was powered on. Try closing the door and rebooting the reader to correct this. Also, check that the shrouds are installed and properly fastened, as electrical noise penetrating the measurement chamber can also cause this error.
Why am I getting error 0700 on my Biotek ELx808IU Laboratory Equipment?
Error 0700 can be caused by electrical noise penetrating the measurement chamber, ambient light leak, or a missing/loose filter. Check that the shrouds are installed and properly fastened, the plate chamber door is closed, and the filter wheel doesn’t have empty positions.
Why does my Biotek Laboratory Equipment show error 0800?
Error 0800 can be caused by electrical noise, ambient light, or filter issues. Ensure shrouds are fastened, the chamber door is closed, and the filter wheel has no empty positions.
What causes error 0300 0301 on Biotek Laboratory Equipment?
Error 0300 0301 can be caused by a loose belt or a loose motor pulley, which may cause the carrier to ignore movement instructions. Check the carrier: the lamp may be out or blocked by a faulty filter wheel. Also, the filter wheel might be missing or loose, or a loose filter may be impeding its movement, so check the filter wheel.
What to do if Biotek Laboratory Equipment shows error 0C00?
Error 0C00 indicates a printer issue. Ensure the printer is connected, online, and has paper.
What causes error 0400 0401 on Biotek ELx808IU Laboratory Equipment?
Error 0400 0401 can be due to a mechanical obstruction, a slipping carrier belt or pulley, a loose filter impeding filter wheel movement, or a defective motor drive circuit/12-volt power problem. Check for obstructions, the carrier belt or pulley, the filter wheel, and the motor drive circuit and power supply.
How to troubleshoot error 0200 0201 on Biotek ELx808IU?
Error 0200 0201 can be caused by a missing or loose filter wheel, a loose filter impeding filter wheel movement, or limited carrier movement. Check the carrier path for obstructions. The wheel is shipped in a separate envelope and may not be installed or may have broken free of the magnetic hub. To confirm if a loose filter is impeding the movement of the wheel, remove the wheel.
What does error 0500 mean on Biotek Laboratory Equipment?
Error 0500 to 0506 indicates that all filter locations must have either a filter or a filter blank installed. Ensure that the filter wheel is installed properly.
Brand | Biotek |
---|---|
Model | ELx808IU |
Category | Laboratory Equipment |
Language | English |
Describes the reader's capabilities, models, and basic features.
Instructs on powering up and verifying instrument functionality.
Details procedures to ensure the instrument operates correctly before use.
Introduces using the Gen5 software for reader control and data analysis.
Explains the process of performing a microplate read.
Details various tests like System Test, Absorbance Plate Test.
Provides step-by-step instructions for executing qualification tests.
Lists essential safety advice before performing maintenance.