How to resolve error 1000 on Biotek ELx808IU?
- Ccindy36Sep 9, 2025
To resolve error 1000, re-download software basecode and assays.
How to resolve error 1000 on Biotek ELx808IU?
To resolve error 1000, re-download software basecode and assays.
What does error 0A00 mean on Biotek ELx808IU?
Error 0A00, 0A01 to 0A05 indicates that the lamp has failed since power-up. Check that the lamp is lit and replace it if necessary.
What to do if I get error 0900 on my Biotek ELx808IU?
Error 0900 can occur if the read chamber door has opened or closed relative to when the unit was powered on. Try closing the door and rebooting the reader to correct this. Also, check that the shrouds are installed and properly fastened, as electrical noise penetrating the measurement chamber can also cause this error.
Why am I getting error 0700 on my Biotek ELx808IU Laboratory Equipment?
Error 0700 can be caused by electrical noise penetrating the measurement chamber, ambient light leak, or a missing/loose filter. Check that the shrouds are installed and properly fastened, the plate chamber door is closed, and the filter wheel doesn’t have empty positions.
Why does my Biotek Laboratory Equipment show error 0800?
Error 0800 can be caused by electrical noise, ambient light, or filter issues. Ensure shrouds are fastened, the chamber door is closed, and the filter wheel has no empty positions.
What causes error 0300 0301 on Biotek Laboratory Equipment?
Error 0300 0301 can be caused by a loose belt or a loose motor pulley, which may cause the carrier to ignore movement instructions. Check the carrier: the lamp may be out or blocked by a faulty filter wheel. Also, the filter wheel might be missing or loose, or a loose filter may be impeding its movement, so check the filter wheel.
What to do if Biotek Laboratory Equipment displays error A900?
Power instrument off, then back on.
What to do if the Biotek Laboratory Equipment lamp is out?
If the lamp is out, check that the lamp is lit when the unit is ON and replace the lamp if necessary. Also, ensure that the filters in the wheel match the filters programmed in the software, in case of a missing, defective, or misplaced filter or filter blank.
What does error 0D00 mean on Biotek ELx808IU?
Error 0D00 indicates internal software corruption. Re-download software basecode and assays.
What does error 0500 mean on Biotek Laboratory Equipment?
Error 0500 to 0506 indicates that all filter locations must have either a filter or a filter blank installed. Ensure that the filter wheel is installed properly.
Describes the reader's capabilities, models, and basic features.
Instructs on powering up and verifying instrument functionality.
Details procedures to ensure the instrument operates correctly before use.
Introduces using the Gen5 software for reader control and data analysis.
Explains the process of performing a microplate read.
Details various tests like System Test, Absorbance Plate Test.
Provides step-by-step instructions for executing qualification tests.