the battery) in ampere*hours (A*hrs) of the battery is calculated and displayed. To run a battery test,
follow these steps:
Key presses Display
I-set, enter current value with
numerical keys, then press
Enter.
Set the DC Load to constant current mode and set the
current you want to draw from the battery. (You may have
to press the I-set key twice to be able to set the current
value.)
Shift + Battery MIN VOLT= 0.10V This indicates you're being asked for the
voltage level where the battery test will be over.
Enter voltage level and press
Enter.
When you press Enter, the test will start. The CC
annunciator will be on.
or Shows the power level and the ampere*hours (A*hrs)
accumulated so far.
When the battery voltage drops below the specified value, the test will turn off and you will see
the OFF annunciator.
or Display the total charge delivered by the battery in
ampere*hours (A*hrs).
Shift + Battery Turns battery testing mode off.
NOTE: you must make sure that you press Shift + Battery at the end of the test. Otherwise, the
instrument will not respond to any key presses except or .
Transient operation
The transient condition allows switching between two different load values. There are three different
types of transient operation.
Continuous transient operation
In continuous transient operation, the load is continuously switched between two load values. An
example is shown in the following figure:
Note that this transient behavior works with any of the operation modes CC, CV, CW, or CR. Here
are the keystrokes used to setup this continuous transient condition:
Keys Display
I-set Standard display of voltage and current (or, you'll be asked to enter a
current value).
8500 DC Load Series Version: February 4, 2009 Page 33 of 76

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 8500 and is the answer not in the manual?

BK Precision 8500 Specifications

General IconGeneral
Maximum Power150 W
Current Range0-30 A
Load ModesCC, CV, CR, CP
Input Voltage100-240 VAC, 50/60 Hz
Voltage Resolution1 mV
Current Resolution1 mA
Ripple and Noise (CV)≤ 10 mVrms
Operating Temperature0°C to 40°C
Storage Temperature-10°C to 70°C
InterfaceRS232
DisplayVFD
Voltage Accuracy0.05% + 5 mV
Current Accuracy0.1% + 10 mA
Ripple and Noise (CC)10 mA rms