14
Test Frequency
Default testing frequency is 1KHz. Push
FREQ
key to select the
desired test frequency.
L/C/R Function Selector
Simply press the
L/C/R
pushbutton to select the desired L, C or R
function.
Relative
Press the RELkey to enter the relative mode and stores the display
reading as a reference value. It will then display all subsequent readings
in relative to reference value. Press the button again to exit the relative
mode.
Notes:
1.
The relative mode can’t be activated if the display value is either
OL
or
0000
”.
2.
Relative mode is only available in manual ranging; however,
activation while in auto-ranging will automatically set the meter to
manual ranging and cause calibration prompts to be displayed in
the recommended ranges.
3.
The relative mode cannot be activated if the meter is set at auto-
ranging with data hold activated.
Tolerance
There are 1%, 5%, 10% and 20% tolerance range. To enter this
tolerance mode, insert the appropriate component as a standard value
into the socket or connect the component to the test probes, then press
the TOLpushbutton to set this value, as the standard reference
tolerance. Similarly, any value which appears on the LCD display, such
as DH or MAX/MIN/AVG, can be used as a standard value to sort
components. Press this button again to cycle through 1%, 5%, 10% and
20% tolerance as desire.
This function is designed for convenient component sorting. An audible
tone will sound whenever the component under test exceeds the setting

Questions and Answers: