EasyManua.ls Logo

BMC RESmart BPAP - Page 15

BMC RESmart BPAP
21 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual for RESmart
®
BPAP V1.1
12/18
11
1. Connect the air hole of the
pressure sensor YL on the main
board with the bottom hole(hole#1)
in the front of the blower outlet with
a canal of 12cm in length and 2mm
in inner-diameter.
2. Crosswise connect the two air
holes of the pressure sensor LL on
the main board with the two top
holes (hole#2 & hole#3) around the
blower outlet via two canals of 8cm
in length and 2mm in
inner-diameter.
12
Clean up the blower wires and wind
them with a tube. Insert the plugs
into sockets PM1 and PM2. Put the
lead of thermistor into socket PRT
on main board. Connect the main
board to 24V power supply.
Blower cable
to PM2
24V power
supply
Thermistor lead
to PRT
Blower driving
lead to PM1
Blower wire
and winding
tube
13
Draw the shield back backwards
and put the 9-pin plug into the
corresponding socket on its back.
Lay the main board onto the 4
supporting poles on the shield
bottom and set it into the power
board via the fixing pole on PCB
and get the main board fixed.
Put blower wires on the side into the
gap between blower unit and shield
bottom.