Cmd Meaning PC->964 964->PC
10 Pressure?
= float * Multiplier
float: Pressure always in bar
byte Multiplier:
0: 1.0 (Unit bar)
1: 14.50778 (Unit psi)
2: 750.0617 (Unit Torr)
3: 0.1 (Unit MPa)
*10# *10#float,byte
11 Temperature?
float: in Kelvin
*11# *11#float
12 Operating hours? *12# *12#long
13 High Alarm Parameter?
(first retrieve concentration unit)
float: Limit in present unit
byte1: Enabled
byte2: Latching
*13# *13#float,byte1, byte2
14 Low Alarm Parameter?
(first retrieve concentration unit)
float: Limit in present unit
byte1: Enabled
byte2: Latching
*14# *14#float,byte1,
byte2
15 Set High Alarm Limit (High Limit > Low Limit!)
(first retrieve concentration unit)
*15#float
*15#
16 Set Low Alarm Limit (High Limit > Low Limit!)
(first retrieve concentration unit)
*16#float *16#
17 Set High Alarm Latching
byte 1: Latching 0: not Latching
*17#byte *17#
18 Set Low Alarm Latching
byte 1: Latching 0: not Latching
*18#byte *18#
19 Set High Alarm Enabled
byte 1: Enabled 0: not Enabled
*19#byte *19#
20 Set Low Alarm Enabled
byte 1: Enabled 0: not Enabled
*20#byte *20#
21 Normalizing temperature?
float: in Kelvin
*21# *21#float
23 Normalizing pressure?
= float x Multiplier
float: Pressure always in bar
byte Multiplier:
0: 1.0 (Unit bar)
1: 14.50778 (Unit psi)
2: 750.0617 (Unit Torr)
3: 0.1 (Unit MPa)
*23# *23#float,byte
25 Molecular weight of carrier gas?
byte:
0: Oxygen : 31.9988 g/mol
1: Air: 29.0 g/mol
*25# *25#byte
29 Time?
hh,mm,ss
*29# *29#byte,byte,
byte