Cmd Meaning PC->964 964->PC
30 Set Hour *30#byte *30#
31 Set Minute *31#byte *31#
32 Set Second *32#byte *32#
33 Date Format?
0: DD.MM.YY
1: MM/DD/YY
*33# *33#byte
34 Set Date Format
0: DD.MM.YY
1: MM/DD/YY
*34#byte *34#
35 Date?
DD,MM,YY
*35# *35#byte,byte,byte
36 Set Day *36#byte *36#
37 Set Month *37#byte *37#
38 Set Year (YY) *38#byte *38#
39 RS-232 Timed/Polled?
1: Timed
0: Polled
*39# *39#byte
40 Set RS-232 Timed/Polled
1: Timed
0: Polled
*40#byte *40#
41 RS-232-Interval? (if Timed)
byte. Seconds
*41# *41#byte
42 Set RS-232-Interval (wenn Timed)
byte. Seconds
*42#byte *42#
43 Set Analog Output (Simulation)
Range 0.0: 0 V/ 4mA... 1.0: 10V /20 mA
2.0: normal operation
*43#float *43#
44 Autozero Interval?
byte: hours
wenn 0: no Autozero
*44# *44#byte
45 Set Autozero Interval
byte: hours
if 0: no Autozero
*45#byte *45#
46 Alarm Beep?
1: On
0: Off
*46# *46#byte
47 Set Alarm Beep
1: On
0: Off
*47#byte *47#
48 Cuvette Status?
float: % Dirty
0: Clean
*48# *48#float
49 Send Error Log (max. 16 Entries)
Format: YY,MM,DD,HH,MM,SS,Error
Error has the same format as described in table on page 21
*49# *49#byte1,
byte2,byte3,
byte4,byte5,
byte6,word#...