OZONE ANALYZER BMT 964 Manual, Rev. 04/2021
36
Cmd Meaning PC->964 964->PC
50 Send Event Log (max. 48 Entries)
Format: YY,MM, DD,HH,MM,SS, byte 7 /additional float:
*50# *50#byte1,
byte2,byte3,
byte4,byte5,
byte6,byte7, float #
byte1,...
7-Bits Meaning float Parameter
0 Switched on pressure [bar]
1 switched off Temperature when switched
on [K]
2 Zeroed Dirty (0 – 100 %)
3 High Alarm High Alarm Limit
4 Low Alarm Low Alarm Limit
83 Trigger Zero (Parameter for security)
float: Dirty
If Autozero Interval > 0 ist, this command will take about 20 s
*83#3.14159 *83#float
85 Firmware-Version? *85# *85#float
86 Status (decimally coded, as decribed in table on page 21): *86# *86#word
91 Set Link-Mode Timeout (after switching on always 10 s)
byte: Seconds
*91#byte *91#
93 Alarm Relays closing on Alarm?
1: closing
0: opening
*93# *93#byte
94 Set Alarm Relay Mode
1: closing
0: opening
*94#byte *94#
95 Set User Baud Rate
0: 2400
1: 4800
2: 9600
3: 19200
4: 38400
*95#byte *95#
98 Reset all parameters to standard factory setting *98# *98#
99 Set PIN (prevents changing parameters)
0: no PIN protection
*99#word
(0..9999)
*99#
160 Set Purge Time
byte: Seconds (10..100s)
*160#byte *160#
161 Get Purge Time
byte: Seconds
*161# *161#byte