ACS 553, ACS 563 Service Manual
1 689 975 295 2021-04-13Robert Bosch GmbH
37
Diagnostics and Testing
The Keypad Test is used to determine if the
keys on the control panel are functioning when
in the table below:
• F1 • 1 • 6 • Up Arrow
• F2 • 2 • 7 • Down Arrow
• F3 • 3 • Left Arrow
• F4 • 4 • 9 • Right Arrow
• F5 • 5 • 0 -----
1. From the Service Menu, scroll to
and select icon. The following
displays:
The X displayed represents the key to be
scroll up automatically.
This test cannot be aborted. Once complete, the
display will automatically return to the service
menu.
Setting the Boot Mode makes it possible to boot into
a mode other than Normal (default). The choices
to boot are Normal, Filter Change Required, Leak
Test Required, and End-User Install.
1. From the Service Menu, scroll to
and select icon.
2. icon.
If the password is valid, the following displays:
3. Scroll to the boot mode desired and select
icon. The unit will boot to the selected
4. To leave set boot mode without change, select
icon to return to the Service Menu.
Unless absolutely necessary, resetting the
board is not recommended. Resetting the board
will erase any calibration/customization and
restore the factory defaults.
1. From the Service Menu, scroll to
and select icon. The following
displays:
2. To reset the board to the default values,
select icon.
3.
icon to return to the Service Menu.