EasyManua.ls Logo

Bosch ETT 008.31 - Fine-Tuning the Indicating Accuracy

Bosch ETT 008.31
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
6.3 Fine-tuning the indicating accuracy
In devices for which calibration is compulsory by law, the calibration seal on the rear wall of the device must be removed for fine-tuning.
To do so, loosen the screw beneath the seal and swivel the safety panel to the side (function switch S 60 is now accessible).
The gas outlet must lead to the open air and have a sufficiently wide cross section. If the required hose length exceeds 2 m, the hose
diameter must be increased. Neither pressure nor vacuum must be allowed to form in the hose.
Particularly during fine-tuning, the zero-gas inlet must be able to take in perfectly clean outside air. In case of doubt, bring outside air
in from a suitable atmosphere using a hose.
Test-gas connection
The test gases which are required one after the other for the purpose of fine-tuning must be connected up to the calibration-gas inlet.
Set an admission pressure of approx. 0.7 bar at the bottle-pressure reducer. At this pressure, the flow resistance of the integrated non-
return valve is overcome and a suitable flow rate achieved (up to 4 l/min). If a flow meter is connected to the gas outlet, the flow can
be adjusted (to a minimum of 0.5 l/min).
Correct sequence for fine-tuning
The following order of procedure must be complied with:
Measuring channel Fine-tuning with test gas
1. HC measuring channel, Upper-range value Nr. 5, 3.800 ppm C
3
H
8
small measuring range Mean value Nr. 4, 2.000 ppm C
3
H
8
2. CO measuring channel, Upper-range value Nr. 6, 1% CO
small measuring range Mean value
3. CO
2
measuring channel Upper-range value Nr. 1, 14% CO
2
Mean value Nr. 4, 7.5 % CO
2
4. HC measuring channel, Upper-range value Nr. 2, 16.000 ppm C
3
H
8
large measuring range Mean value Nr. 6, 8.000 ppm C
3
H
8
5. CO measuring channel, Upper-range value Nr. 3, 9% CO
large measuring range Mean value Nr. 4, 5% CO
Within a measuring range,
proceed as follows (using the example of No. 1 HC channel, small measuring range, engine-speed LED does not light up):
a) Fine-tuning the upper-range value
Set function switch S 60 to position 2 (upper-range values). Select the HC channel using the "Select" button, and simultaneously
press "Enter o" to switch on the fine-tuning function.
°C display window:
E. EiE. Ei
E. EiE. Ei
E. Ei (set upper-range value)
HC display window: Current measured value
Engine-speed window: The calibration factor accepted during the last fine-tuning is displayed.
Introduce a flow of test gas no. 5 into the device via the calibration-gas inlet.
Change the calibration factor indicated in the engine-speed display window with the "ENTER o" or "TEST u" buttons in such a
way that the value currently displayed in the HC window exactly matches the certified test-gas value (this value is displayed with
a certain delay after the calibration factor is changed).
! Do not press any buttons!
Introduce a flow of mean-value gas (test gas no. 4) into the analyzer and check that the display deviates from the test-gas value
by less than 2%. In this case, the mean value does not have to be adjusted.
If the display differs from the test-gas value by more than 2%, the mean value must be re-calibrated as described in b) below.
In all cases, now press the "Select" button. The calibration factor of the upper-range value is now accepted and the next possible
function displayed.

Table of Contents

Related product manuals