EasyManua.ls Logo

Bosch ETT 008.31 - Faulty Oil-Temperature Measuring Channel; Faulty O 2 Measuring Channel

Bosch ETT 008.31
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
8.11 Faulty oil-temperature measuring channel
- Check the sensor resistance:
Pin1 to pin 2 = 1005
+ 5 @ 25 °C
1530
+12 @ 90 °C
or check whether the fault can be remedied with a new sensor.
- Sensor OK: switch off and open the exhaust-gas analyzer.
Check the connector for continuity to the sensor board, measure the resistance. Refer to Section 8.15, "Checking the sensor board"
for the location of test points.
Check: - Connector pin 1 to X 18 (or to C43)
- Connector pin 2 to X 83, MA
- Connector pin 2 to X 16 on motherboard, MA
- Connector pin 3 to shielding panel on sensor board.
Volume resistance during all measurements < 100mΩ. If this criterion is not met, replace the sensor board as described in Section 10.9.
- Switch on the analyzer
- With the digital multimeter (DMM), check on the sensor board:
X4 (P5VREF) to X83 (MA): UDC = 5V
+ 10 mV
X 18 (or connector pin 1) to X 83 (MA): UDC = 4.60 to 4.70V.
Depending on U
REF
perform tests on/fine-tune the motherboard to Section 8.17, replace the connector to the motherboard or replace
the sensor board to Section 10.9.
8.12 Faulty O
2
measuring channel
Error messages which indicate a fault in this channel:
0FF50FF5
0FF50FF5
0FF5 During calibration, the zero window is exceeded at the O
2
amplifier output (- 10 to + 80 mV). This indicates oxygen in the
calibration gas, faulty O
2
measuring sensor or faulty O
2
amplifier.
EnPFEnPF
EnPFEnPF
EnPF When air is flowing through the exhaust-gas analyzer (system adjustment = adjustment of sensitivity of O
2
measuring channel),
a voltage threshold of 1.36 V is not reached at the O
2
amplifier output. Faulty O
2
measuring sensor or faulty O
2
amplifier.
u u
u u
u - - - O
2
amplifier signal is negative (< - 10 mV), probe connector has reversed polarity or amplifier is faulty.
If the error message "
0FF50FF5
0FF50FF5
0FF5" (zero fault) or "
EnPFEnPF
EnPFEnPF
EnPF" (sensitivity fault) appear, a new O
2
measuring sensor should always be installed
before performing any further tests. After this, the upper-range value and zero of the O
2
measuring channel must be re-defined.
Upper-range value: Press pump button, wait until system adjustment is complete.
Zero adjustment: Perform to Section 4.1.8.
If error messages still occur, open the exhaust-gas analyzer and check the O2 measuring channel.
Connect the digital multimeter to X 53 of the sensor board (reference point X 83, MA). Refer to diagram in Section 8.15, "Checking the
sensor board".
Press the pump button (system adjustment starts, air flows through the device).
Measured value at DMM (X 53) = UDC 1.36 V to 5.4 V. This measured value is achieved if the O
2
measuring sensor and amplifier are
not faulty.
Check the O
2
signal amplifier:
- Connect the DC calibrator (0...15 mV) to amplifier input X 51 (reference point MA = X 51)
- Set calibrator to 0 mV. Measured value at DMM = 0 V + 20 mV
- Set calibrator to + 15 mV. Measured value at DMM = 4.1 V + 0.2 V.
ZFor the purpose of testing, the zero and sensitivity of the O2 measuring channel can also be adjusted with the above two settings.
To do so:
- Apply + 15 mV, press the pump button, wait until system adjustment is complete
- Apply 0 mV, start zero adjustment in input mode (see Section 3.1.8).
In analysis mode (see Section 4.1.10), each intermediate value can be read on the display in O
2
% vol.
Replace defective components as appropriate.

Table of Contents

Related product manuals