54
9.2.4 Entering the date/time
Sequence Keypad entry Exhaust-gas analyzer Feedback
1. D Receives and sends back D
2. CR Receives and starts input procedure Day:
by transmitting "TAG"
3. 1(2) digits, CR Receives and stores both digits as MONTH:
DAY in the clock element.
Transmits "MONTH"
4. 1(2) digits, CR Receives and stroes both digits as YEAR:
MONTH in the clock element.
Transmits "YEAR"
5. 1(2) digits, CR Receives and stores both digits as Hour:
YEAR in the clock element.
Transmits "HOUR"
6. 1(2) digits, CR Receives and stores both digits as MINUTE:
HOUR in the clock element.
Transmits "MINUTE"
7. 1(2) digits, CR Receives and stores both digits as MINUTE
in the clock element. End of input
of DATUM/UHRZEIT
i The device monitors the plausibility of the entry. It interrupts an obviously incorrect entry (e.g. Date: 33. Month: 13 etc.)
with a renewed prompt (e.g. "*Month:") to enter the date/time again.
8. CR End of input
i If you are only changing individual values, you do not have to enter digits for unchanged values.