2-12
MRI Scan Procedure Protocol
1. Programming the Pulse Generator for a Scan
[1] If MRI Protection Time-out is displayed as “Off”, the pulse generator remains in MRI Protection Mode until manually reprogrammed; [2]
Twenty-four hour time format is used; [3] Column indicates date measurement was taken
Figure 2–8. Sample settings report and checklist printout
The End Session button will end the current programmer session with the pulse generator
remaining in MRI Protection Mode (Figure 2–9 End Session Confirmation dialog on page 2-13).