C-1
IMAGEREADY PACING SYSTEM COMPONENTS FOR 1.5 T AND 3 T
APPENDIX C
Only specific combinations of pulse generators and leads constitute an ImageReady Pacing
System. Consult the following tables to determine which combinations are valid for use with only
1.5 T scanners and which are valid for use with both 1.5 T and 3 T scanners.
Table C–1. Valid Combinations of Pulse Generators and Leads to Use in 1.5 T and 3 T Environments
INGEVITY MRI Leads
only
FINELINE II Leads only
Combination of one
INGEVITY MRI Lead and one
FINELINE II Lead
ESSENTIO MRI Pulse
Generator
PROPONENT MRI
Pulse Generator
ACCOLADE MRI Pulse
Generator
1.5 T or 3 T scanner
allowed.
Normal Operating Mode
or First Level Controlled
Operating Mode.
1.5 T scanner only.
3 T scanner not allowed.
Normal Operating Mode
only.
1.5 T scanner only.
3 T scanner not allowed.
Normal Operating Mode only.
Table C–2. ImageReady MR Conditional Pacing System Components for 1.5 T and 3 T
Component Model Number(s) MR Status Valid
Combinations
Pulse Generators
For valid
combinations for 1.5
T and 3 T see the
“Valid Combinations
of Pulse Generators
and Leads to Use in
1.5 T and 3 T
Environments” table
above.
ESSENTIO MRI L110, L111, L131 MR Conditional
PROPONENT MRI L210, L211, L231 MR Conditional
ACCOLADE MRI L310, L311, L331 MR Conditional
Leads and Accessories
FINELINE II Sterox Pacing Lead 4456, 4457, 4458,
4459, 4479, 4480
MR Conditional
FINELINE II Sterox EZ Pacing Lead 4469, 4470, 4471,
4472, 4473, 4474
MR Conditional
Suture Sleeve for FINELINE II leads 6220, 6221 MR Conditional
INGEVITY MRI Pacing Lead (Tined Fixation) 7731, 7732, 7735,
7736
MR Conditional
INGEVITY MRI Pacing Lead (Extendable/
Retractable Fixation)
7740, 7741, 7742 MR Conditional
Suture Sleeve for INGEVITY MRI leads 6402 MR Conditional
IS-1 Lead Port Plug 7145 MR Conditional
ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) and PRM Software Application
ZOOM LATITUDE PRM 3120 MR Unsafe
a
ZOOM LATITUDE PRM Software App. 2869 N/A
a. See PRM is MR Unsafe Warning regarding the PRM.