Introduction to MR Conditional Pacing
System Configuration for 1.5 T
1-3
NOTE: 3 T MRI scanners should be operated in Circularly Polarized mode or Dual Channel
Multi-Transmission mode since ImageReady Pacing Systems have been evaluated against 3 T
MRI scanners during Circularly Polarized and Dual Channel Multi-Transmission RF operation
modes. Multi-Transmission RF operation with greater than two transmit channels has not been
evaluated.
System Configuration for 1.5 T
Table 1–2. System Configuration for 1.5 T
Component Model Number(s) MR Status
Pulse Generators
ESSENTIO MRI Pulse Generator L110, L111, L131 MR Conditional
PROPONENT MRI Pulse Generator L210, L211, L231 MR Conditional
ACCOLADE MRI Pulse Generator L310, L311, L331 MR Conditional
Leads and Accessories
FINELINE II Sterox / Sterox EZ Leads
FINELINE II Sterox Pacing Lead 4456, 4457, 4458, 4459, 4479, 4480 MR Conditional
FINELINE II Sterox EZ Pacing Lead 4469, 4470, 4471, 4472, 4473, 4474 MR Conditional
Suture Sleeve for FINELINE II leads 6220, 6221 MR Conditional
IS-1 Lead Port Plug 7145 MR Conditional
INGEVITY MRI Leads
INGEVITY MRI Pacing Lead (Tined Fixation) 7731, 7732, 7735, 7736 MR Conditional
INGEVITY MRI Pacing Lead (Extendable/
Retractable Fixation)
7740, 7741, 7742 MR Conditional
Suture Sleeve for INGEVITY MRI leads 6402 MR Conditional
IS-1 Lead Port Plug 7145 MR Conditional
ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) and PRM Software Application
ZOOM LATITUDE PRM 3120 MR Unsafe
a
ZOOM LATITUDE PRM Software App. 2869 N/A
a. See PRM is MR Unsafe Warning regarding the PRM.
System Configuration for 3 T
Table 1–3. System Configuration for 3 T
Component Model Number(s) MR Status
Pulse Generators
ESSENTIO MRI Pulse Generator L110, L111, L131 MR Conditional
PROPONENT MRI Pulse Generator L210, L211, L231 MR Conditional
ACCOLADE MRI Pulse Generator L310, L311, L331 MR Conditional
Leads and Accessories
INGEVITY MRI Leads
INGEVITY MRI Pacing Lead (Tined Fixation) 7731, 7732, 7735, 7736 MR Conditional
INGEVITY MRI Pacing Lead (Extendable/
Retractable Fixation)
7740, 7741, 7742 MR Conditional
Suture Sleeve for INGEVITY MRI leads 6402 MR Conditional