Vercise DBS Remote Control 4 Directions for Use
Vercise™ DBS Remote Control 4 Directions for Use
92378769-02 38 of 1157
en
Charge Stimulator Now or Disable MRI Mode Screen (Rechargeable Stimulators Only)
If MRI Mode has already been enabled and the Stimulator battery power falls below the recommended
value, the Remote Control will display a message instructing the patient to charge the Stimulator.
To charge the Stimulator without disabling MRI Mode:
1. Do not press
.
2. Instruct the patient to charge the Stimulator.
3. Check the Remote Control to conrm that the error message has cleared.
4. Navigate to the Home screen on the Remote Control by pressing
on the side panel of the Remote
Control and conrm that the MR Conditional Symbol
is displayed on the Home screen.
The patient can also disable the MRI Mode before charging the Stimulator:
1. Press
to disable the MRI Mode.
2. Instruct the patient to fully charge the Stimulator.
3. Check the Remote Control to conrm that the error message has cleared.
4. Enable MRI Mode by following instructions in the “Enabling MRI Mode” section of this manual.
Caution: The Charger and Remote Control are MR Unsafe and must not be brought into the MRI
scanner room.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Boston Scientific Vercise DB-5572-1A and is the answer not in the manual?

Boston Scientific Vercise DB-5572-1A Specifications

General IconGeneral
BrandBoston Scientific
ModelVercise DB-5572-1A
CategoryMedical Equipment
LanguageEnglish

Related product manuals