7.4 Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
7.5 Wipe disinfection for electrical devices without sterilization
RT: Room temperature
Phase I
► Remove any visible residues with a disposable disinfectant wipe.
► Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
► Observe the specified application time (1 min minimum).
► Clean the contacts in the charging bay with isopropanol or ethyl alco-
hol and a cotton swab. Do not use corrosive chemicals.
DANGER
Risk of electric shock and fire hazard!
► Unplug the power cord before cleaning.
► Do not use flammable or explosive cleaning or
disinfecting solutions.
► Ensure that no fluids will penetrate the product.
CAUTION
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
► Only clean and disinfect the product manually.
► Do not sterilize the product under any circum-
stances.
CAUTION
Risk of damage to the product due to inappropriate
cleaning/disinfecting agents!
► Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer’s
instructions for the respective cleaning/disin-
fecting agent.
► Do not introduce any fluid into the charging
bays.
Phase Step T
[°C/°F]
t
[min]
Conc.
[%]
Water quality Chemicals
IWipe disinfectionRT ≥1 - - Meliseptol Wipes sensitive