8
Aesculap
®
Acculan 4
Acculan 4 Charger GA320
7.4 Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
7.5 Wipe disinfection for electrical devices without sterilization
RT: Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
Clean the contacts in the charging bay with isopropanol or ethyl alco-
hol and a cotton swab. Do not use corrosive chemicals.
DANGER
Risk of electric shock and fire hazard!
Unplug the power cord before cleaning.
Do not use flammable or explosive cleaning or
disinfecting solutions.
Ensure that no fluids will penetrate the product.
CAUTION
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
Only clean and disinfect the product manually.
Do not sterilize the product under any circum-
stances.
CAUTION
Risk of damage to the product due to inappropriate
cleaning/disinfecting agents!
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer’s
instructions for the respective cleaning/disin-
fecting agent.
Do not introduce any fluid into the charging
bays.
Phase Step T
[°C/°F]
t
[min]
Conc.
[%]
Water quality Chemicals
IWipe disinfectionRT 1 - - Meliseptol Wipes sensitive

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Aesculap GA320 and is the answer not in the manual?

Braun Aesculap GA320 Specifications

General IconGeneral
BrandBraun
ModelAesculap GA320
CategoryBattery Charger
LanguageEnglish

Related product manuals