10. Replacing fuses
Mandatory fuse set: 2x – T 1.25 AH, 250 V
► Use a small screwdriver to release the clip on the fuse holder.
► Pull out the fuse holder.
► Replace both fuse sets.
► Reinstall the fuse holder so that it audibly clicks into place.
Note
If the fuses burn out frequently, the device is faulty and should be repaired,
see Technical Service.
11. Technical Service
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
► For service and repairs, please contact your national
B. Braun/Aesculapagency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone: +49 (7461) 95 -1601
Fax: +49 (7461) 14 -939
E-Mail: ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone: +1 (800) 214 -3392
Fax: +1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
12. Accessories/Spare parts
13. Technical data
13.1 Performance data, information about standards
Acculan 4 Charger GA320
DANGER
Risk of death by electric shock!
► Pull the mains plug before replacing the fuse
sets.
DANGER
Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
► Do not modify the product.
Art. no. Approvals Color Length
[m]
TE780 Europe Black 1.5
TE730 Europe Black 5.0
TE734 Great Britain Black 5.0
TE735 USA, Canada, Japan Gray 3.5
Art. no. Designation
TA020112 Fuse link T 1.25 AH, 250 V
TA014534 Instructions for use of Acculan 4 Charger GA320
(A4 for ring binder)
Classification according to
Directive 93/42/EWG
I
Protection class according to
IEC/DINEN60601-1
I
Housing protection rating
according to
IEC/DINEN60529
IP21
For charging Aesculap bat-
tery types
GA346 and GA666, GA676
Mains voltage ranges
(Current intake)
100 V~ to 120 V~ (0.8 A to 0.7 A) /
220 V~ to 240 V~ (0.5 A to 0.4 A)
Frequency 50 Hz to 60 Hz
Charging voltage/Output
voltage
max. 15 V
Charging current/Output
current
max. 1.2 A ± 0.3 A
Operational mode Continuous operation
Device protection T 1.25 AH, 250 V
Form factor: 5 mm x 20 mm
Weight without battery approx. 2.3 kg