6
Aesculap
®
Acculan 4
Acculan 4 Charger GA320
6. Working with the device
6.1 System set-up
Connecting to the power supply
Carry out a visual inspection before connecting the product to the
mains power:
Check the mains power cord for any damage.
Check the charger for any damage (e.g. bent, short-circuited con-
tacts of the charging bays).
Do not use the product if it is damaged, defective or dirty.
Set aside the product if it is damaged.
Clean product if dirty, see Validated reprocessing procedure.
Plug the mains power cord into the device connector 7.
Plug in the mains plug at the mains power socket.
After being connected to the mains power, the charger performs a self-
test. During this time the "Power On” 3 indicator and the
indicator fields for charging bays 2 must be illuminated on the front
panel of the device.
Carry out a visual inspection of the "Power On" indicator 3 and the
indicator fields for charging bays 1.
The indicator fields for the charging bays 2 go out and only
the "Power On" 3 indicator remains lit.
The charger is ready.
Note
In case of any deviation from the sequence described above, see Trouble-
shooting list
Putting out of operation
Note
The Acculan 4 Charger GA320 does not have a power ON/OFF switch. The
all-pole disconnection of the product from the main power supply is only
guaranteed when the power cord/mains connection line is unplugged.
Unplug the power cord from the mains power socket.
Unplug power cord from the device connector 7.
6.2 Safe operation
As excessive battery temperatures will prolong the charging time,
allow batteries to cool down prior to charging.
To remove the battery, if necessary use a cloth as the batteries become
heated during the charging process.
DANGER
Risk of death by electric shock!
Connect the product only to a grounded power
supply.
WARNING
Risk of injury and material damage due to inappro-
priate use of batteries!
Only insert batteries that are indicated in the
Intended Use.
Insert batteries in their correct position.
WARNING
Risk of injury and material damage due to damaged
batteries!
Check batteries for any damage before charg-
ing.
Do not use or recharge damaged batteries.
CAUTION
Total battery failure can occur by leaving the bat-
teries in the charger while disconnected from the
power supply!
Always leave charger connected to the power
supply. Indicator lamp for readiness must illumi-
nate.
Always remove batteries from the charger when
it is disconnected from the power supply.
CAUTION
Loss of capacity/performance due to prolonged
storage of discharged batteries!
After a successfully completed charging process,
leave the battery in the operational charger
(trickle charging).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Aesculap GA320 and is the answer not in the manual?

Braun Aesculap GA320 Specifications

General IconGeneral
BrandBraun
ModelAesculap GA320
CategoryBattery Charger
LanguageEnglish

Related product manuals