6
Aesculap
®
Acculan 4
Acculan 4 Charger GA320
6. Working with the device
6.1 System set-up
Connecting to the power supply
► Carry out a visual inspection before connecting the product to the
mains power:
– Check the mains power cord for any damage.
– Check the charger for any damage (e.g. bent, short-circuited con-
tacts of the charging bays).
► Do not use the product if it is damaged, defective or dirty.
► Set aside the product if it is damaged.
► Clean product if dirty, see Validated reprocessing procedure.
► Plug the mains power cord into the device connector 7.
► Plug in the mains plug at the mains power socket.
After being connected to the mains power, the charger performs a self-
test. During this time the "Power On” 3 indicator and the
indicator fields for charging bays 2 must be illuminated on the front
panel of the device.
► Carry out a visual inspection of the "Power On" indicator 3 and the
indicator fields for charging bays 1.
The indicator fields for the charging bays 2 go out and only
the "Power On" 3 indicator remains lit.
The charger is ready.
Note
In case of any deviation from the sequence described above, see Trouble-
shooting list
Putting out of operation
Note
The Acculan 4 Charger GA320 does not have a power ON/OFF switch. The
all-pole disconnection of the product from the main power supply is only
guaranteed when the power cord/mains connection line is unplugged.
► Unplug the power cord from the mains power socket.
► Unplug power cord from the device connector 7.
6.2 Safe operation
► As excessive battery temperatures will prolong the charging time,
allow batteries to cool down prior to charging.
► To remove the battery, if necessary use a cloth as the batteries become
heated during the charging process.
DANGER
Risk of death by electric shock!
► Connect the product only to a grounded power
supply.
WARNING
Risk of injury and material damage due to inappro-
priate use of batteries!
► Only insert batteries that are indicated in the
Intended Use.
► Insert batteries in their correct position.
WARNING
Risk of injury and material damage due to damaged
batteries!
► Check batteries for any damage before charg-
ing.
► Do not use or recharge damaged batteries.
CAUTION
Total battery failure can occur by leaving the bat-
teries in the charger while disconnected from the
power supply!
► Always leave charger connected to the power
supply. Indicator lamp for readiness must illumi-
nate.
► Always remove batteries from the charger when
it is disconnected from the power supply.
CAUTION
Loss of capacity/performance due to prolonged
storage of discharged batteries!
► After a successfully completed charging process,
leave the battery in the operational charger
(trickle charging).