Operating alarms
Remedy failure and restart infusion.
Drop Alarm / Pressure Alarm
Infusion bottle empty?
Roller clamp closed?
Flow? – Close roller clamp.
Infusion STOP. There may not be a continuous
drip. Insert new infusion line, if necessary.
Occlusion? – Insert infusion line free of kinks
and check the free flow (consider bolus).
Condensed drop chamber? Shake to remove.
Drop sensor fitted/connected? Replace drop
sensor, if necessary.
Air Alarm
Air in administration set? – Insert infusion
line correctly. Vent and reset fluid level in drop
chamber.
Standby Alarm
Alarm after pre-selected pause?
Switch to Standby with key below SF.
Then end pause with key below OFF or extend
pause with key below ON.
Battery Pre-Alarm
Battery pre-alarm 30 min before battery is
discharged:
- Alternately rate and AAA.A are displayed,
- battery indicator flashes,
- audible alarm every 9 sec.
Alarm can be cleared by pressing -key. The
alarm continues in short intervals until the bat-
tery is fully discharged.
Battery Alarm:
- Alternately rate and AAA.A are displayed.
- The text display shows:
"battery discharged, connect to mains”.
- battery indicator flashes,
- red alarm indicator is on,
- audible alarm every 4 sec.,
- operating indicator off,
- staff call.
Switch off device.
Connect to mains/power or 12 V DC power
source.
KOR-Alarm (KVO)
Pre-alarm: audible alarm every 9 sec max.
30 min. End of infusion alarm: Permanent
interval signal tone (also via staff call).
Delay of Alarm Tone
When connected to staff call the alarm tone
on the device can be suppressed for 10 min.
(This function can be activated via service only.)
Further Alarms / Displays
"pump door open”? – Close door.
"invalid rate”? – Enter different value.
Device Alarms
When display indicates "device alarm” an
audible alarm sounds permanently.
Press ON/OFF-key repeatedly until display
indication "do not press any key until display is
off”. Pump switches off automatically after a
few seconds.
Switch on device again.
In case of repeated device alarm inform service.
Alarms
13