To Change the Delivery Rate
➤ Press START/STOP.
Green operating indicator goes out.
➤ Press C.
Display reset to 000.0
➤ Enter new delivery rate. (No setting possible,
when C has not been pressed.).
➤ Press START/STOP to restart infusion. (Alarm
if no restart within 2 min).
To Change the Rate Without Interrupting the
Infusion (Function can be deactivated by
Service staff)
➤ Enter new rate.
➤ Press key below RATE. Rate is transferred to
the upper level as large numbers.
➤ Pump operates at new rate. (If new rate is
not confirmed within 10 sec, infusion continues
at previous rate.)
Target Volume (Volume Pre-selection)
The target volume will be administered
independent on the infused volume.
➤ Press key below VOL.
➤ Enter target volume via keypad and confirm
with VOL. Values between 0.1 and 9999.9 ml.
After confirming the display shows the residual
volume instead of the target volume.
Note:
When target- / residual volume has been
administered, the device switches to KOR-mode.
Stop pump and press VOL. Clear displayed
residual volume with C (in display appears
target volume ----,- ml) and confirm with VOL.
If desired, enter new target volume.
To continue infusion, the target volume has to
be displayed as ----,- ml or a new target
volume has to be set.
Target Time (Time Pre-selection)
➤ Press key below TIME.
Enter target time via keypad (50 min = 50;
2 h 30 min = 2 3 0).
➤ Confirm: Press key below TIME again,
instead of the target time the residual time is
displayed now.
➤ Correction
Press C. Display ––h––m. New entry.
Note:
When pre-selected time has been expired, the
device switches to KOR-mode. Stop pump and
press TIME. Clear resid. time with C (in display
appears target time ––h––m) and confirm with
TIME.
If desired, enter new target time. To continue
infusion, the target time has to be displayed as
––h––m or a new target time has to be set.
Rate Calculation
(Displayed delivery rate must be 000.0)
➤ Enter volume and time.
The delivery rate is calculated automatically and
displayed (rounded to one decimal place).
➤ Confirm: Press key below RATE.
➤ Start infusion with START/STOP.
Clear Time/Volume
➤ Press STOP, then key below VOL or TIME.
➤ Press C to clear.
1x: Target Vol/Time = ––––.-ml / ––h--m
2x: Infused Volume/Real Time
= 0.0ml / 00h00m
Additional Settings
7