Infusion
1. Ensure reliable installation
Never position infusion bottle below pump level.
Connect staff call.
Insert spike vertically into infusion bottle.
Fill lower part of drop chamber to max. 2/3.
Open roller clamp.
2. Filling and Venting
Fill infusion line from bottom to top.
Close roller clamp.
3. Switch On with
Green mains/power control or yellow battery
control, alarm tone and display LED’s are briefly
activated.
Displayed type of infusion line must be identical
to the used infusion line. Confirm with “YES”.
IFPE: Change of type infusion line is possible if
function is activated via Service. Only in this
case “NEXT” appears for selection of the type of
infusion line.
4. Insert Infusion Line
Press door opener.
Insert infusion line: First locate in free flow
clamp, then insert vertical to bottom and fit
properly into the air sensor.
Keep infusion line in place at air sensor.
Close door.
Open roller clamp completely. There may not
be a continuous drip.
Place drop sensor on drop chamber
(if necessary, use an appropriate adapter).
5. Puncture
6. Setting the Delivery Rate
In the range: 0.1 - 999.9 ml/h (IFPe in the
range: 0.1 - 600.0 ml/h) and check
(selectable in 0.1 ml/h-increments).
Correction: Press C, then set new rate. Select
additional settings, if desired (see page 9).
7. Press START
Operating symbol appears on the display and
the green operating indicator lights.
8. Stop the infusion
Press STOP. Green operating indicator goes
out
Close roller clamp.
Press door opener. Infusion line will be
clamped off when opening the door.
Remove infusion line after pressing free flow
clamp. - Replacement: Insert new infusion line
as described. Then press START again.
To end switch off
Press for 2 sec.
Operation
6
free flow clamp
air sensor