16
Technical Data
Type of unit Volumetric infusion pump
Classification (acc. to IEC/EN 60601-1) defibrillator-proof, CF equipment
Protection Class I;
IP 22 (Moisture protection: drip protected)
Class (acc. to Directive 93/42 EEC) II b
Rated voltage / current 230 V AC~ (0.06 A), 50/60 Hz
Mains fuse T 0.16 A or
200 V/230 V/240 V AC~ * (0.06 A), 50/60 Hz
Mains fuse T 0.16 A or
100 V/110 V/120 V AC~ * (0.12 A), 50/60 Hz
Mains fuse T 0.315 A
* Mains voltage can be selected at appliance inlet.
External extra-low voltage 12 V DC
Staff call Max. 24 V / 1 A / 24 VA
Arbitrary connection polarity (VDE 0834)
EMC EN 55011
IEC/EN 60601-1-2 and IEC/EN 60601-2-24
Time of operation 100 % (continuous operation)
Operating conditions
-Relative humidity 30 % ... 90 % (without condensation)
-Temperature + 18 °C ... + 35 °C
-Atmospheric pressure 700 mbar ... 1060 mbar
Storage conditions
-Relative humidity 10 % ... 90 %
-Temperature - 25 °C ... + 55 °C
-Atmospheric pressure 500 mbar ... 1060 mbar
Battery type (rechargeable) NiCd (7.2 V; 1.2 Ah)
Operating time of rech. battery > 3.0 h at highest delivery rate
Recharging time > 16 h
Weight / Dimensions (WxHxD) Approx. 3.1 kg / 140 x 240 x 200 mm