Air detector Technical sensitivity
Air bubbles > 0.01 ml
Alarm triggering: With air bubble size
of typ. 0.3 ml
1)
(limit value 0.4 ml) or 1.5 ml/h
2)
(cumulative value of 1 h as of air bubble
volume 0.01 ml)
1)
Can be set from 0.01 to 0.3 ml via service program only
2)
Can be set from 0.5 to 3.5 ml/h via service program only
Accuracy of set delivery rate typ. ± 5 % with Intrafix® P Infusion set,.
measurement time 24 h, temperature 22 °C
Delivery range 0.1 ... 999.9 ml/h (0.1 ml/h-increments)
Software IFPE
0.1 600.0 ml/h (0.1 ml/h-increments)
Software IFPe
Delivery pre-selection 0.1 ... 9999.9 ml (in 0.1 ml-increments)
Occlusion alarm pressures
Alarm reaction time
Max. bolus volume
(measured at 22 °C with Intrafix® P set)
Mechanical occlusion pressure limit Occlusion alarm pressure max. 3.0 bar (300 kPa)
under fault conditions max. bolus volume 2 ml
Alarm in case of incorrect dosage In case of incorrect dosage of max. 0.6 ml due
to apparatus malfunction the pump switches off
automatically.
KOR-rate (KVO) Delivery rate > 10 ml/h = 3 ml/h
Delivery rate < 10 ml/h = 1 ml/h
Delivery rate < 1 ml/h = STOP
17
low high
rate app. 0.6 bar app. 1.2 bar
1 ml/h 17 min 30 min
25 ml/h 45 sec 72 sec
100 ml/h 12 sec 18 sec
0.3 ml 0.5 ml