KOR-Mode (KVO)
After expiring of the pre-selected time or infusi-
on of the pre-selected volume automatic reduc-
tion to keep-open rate (KOR).
KOR and delivery rate flash alternately.
Deactivation and alarm after 30 min.
Switch off with START/STOP.
Status request
Only when infusion is running.
Press 1 x resp. 2 x key below INFO for actual
values.
The display disappears after 10 sec or after all
information has been requested.
Press 3 x key below INFO for battery
capacity in h min and hours of operation.
Additional Settings
8
History Function
In connection with software “IFPE” the
Infusomat® P is equipped with a history func-
tion (memory of events). This permanent
memory records the last 350 events time- and
date related:
- Set delivery rate
- Changes of the rate
- Switching on/off
- Start/Stop of infusion
- Remote Control
- Operating and device alarms
At the moment of an event also the volume
infused from switching on is recorded.
The history function is activated ex works.
After a software-update the function has to be
installed again (service programme: menu 560
calibration history card). With this programme
the function also can be deactivated. A faultless
time-related recording requires exact setting of
time and date.
Recording the data:
The protocol of events only can be transferred
to a computer via interface (MFC interface
lead). For this a terminal programme installed in
the computer has to be selected
(settings: 9600 baud, 1 start – 8 data,
1 stopbit).
For the data transfer the Infusomat® P has to be
switched off and connected to mains.
The protocol can be requested with key “##”
and begins with the latest event. Data are
screened in the terminal programme of the
computer.
To stop data transfer: Press key below END.
The history function mainly is used for failure
analysis for the technical service. A data
transfer via fm system is not possible.
Special Functions