Select Special Mode SF
Depending on the version, several functions may
be deactivated. Contact service.
To set special functions press key below SF
repeatedly until desired special function is
displayed. – Then follow Instructions for Use as
described.
Activate Bolus Function
For additional boli.
Pressing key below ON activates the
function (function is maintained after switching
on again).
T
o change bolus rate:
Press key below RATE. Enter new value and
confirm.
Correction: Press C and re-enter.
Exit bolus function:
Press key below END.
Bolus application during infusion:
Bolus with volume pre-selection
Press key below BOL and release.
Display: BOLUS RELEASE?
Enter bolus. Values between 0.1 and 99.9 ml
(if no entry is made within 10 sec function is
exited automatically).
Press key below YES. Bolus is administered.
After the bolus application the infusion
continues at basic rate.
T
o stop bolus:
Press key below STOP.
Bolus without Volume Pre-selection
Press and hold BOL until a second BOL is
displayed.
Hold both BOL-keys. Bolus is administered
as long as both keys are pressed. Per ml bolus
administered a short audible signal sounds.
Interval bolus
Automatic bolus in set time interval. In case of
manual bolus administration the interval bolus
is skipped.
Select bolus function under SF.
Press key below VOL, TIME or RATE. Enter
values and confirm. After value below TIME has
been confirmed, interval times runs automati-
cally. Exit function with END, set basal rate and
start. Remaining interval time is displayed in
h:min:sec.
A bolus on demand is possible at any time
during infusion at basal rate.
Bolus and dosage calculation
Operation as in volumetric mode.
Depending on the settings in ml (volume),
quantity of active agent (e.g. mg) or in quantity
of active agent per kg weight (mg/kg) the bolus
may be administered. During bolus administrati-
on all three values are displayed.
Standby/Pause
In case of extended interruptions set values are
retained.
Press STOP.
Press key below SF until "Standby” is
displayed.
Press key below ON.
Enter length/duration of interruption interval
or confirm time displayed.
Correction: Press C. Display 00h 00m.
New entry.
Confirm TIME. Timer for interruption on
display is running. Alarm at the end of
interruption interval.
End of interruption interval: Press key below
END.
Special Functions
9