Drug Display
9 selectable drug names can be stored (input via
service program only).
➤ Select with key below (+). Drug is displayed,
also during infusion.
➤ Key below CLR deletes drug name from
display.
➤ Exit selection: Press key below END.
Occlusion Pressure
Due to variable pressure limits shortened alarm
times are possible. Occlusion pressure high/low.
See Technical Data.
➤ Select pressure with key below (+) or (-).
➤ Exit selection: Press key below END (select
pressure as low as possible).
Switch off Drop Control
Caution:
- No alarm will be emitted if the drop control
is switched off and the roller clamp is closed
(underdosage).
- No alarm will be emitted in case of occlusi-
on within the infusion line and pressure sensor
fails.
- Drop control may only be switched off
where underdosage is not critical for the patient
or where the patient is maintained under
constant observation.
Operation without drop control only with volu-
me pre-selection:
➤ Set infusion volume.
(The volume in the infusion bottle has to be suf-
ficient!).
➤ Press key below SF as many times as
necessary for “drop control” to appear.
➤ Press key below OFF. “no drop control”
appears.
➤ To switch on again: Press key below ON.
➤ Exit from selection: Press key below END.
Check Battery Capacity
Remaining battery life time is displayed,
e.g. battery capacity = 02 h 30 min.
In addition the operating hours are displayed.
To switch off display again: Press key below
END.
➤ With device switched off and mains/power
lead detached: Briefly press . Short display
of remaining battery life time after 3 sec. -
Device switched off and mains/power lead
connected: Permanent display of remaining bat-
tery life time.
Battery replacement is recommended if a
remaining life time of less than 2 h is displayed
after 16 h charging time.
Data Lock
Interlocks keys to prevent unauthorised use.
➤ Press key below ON. Keys are interlocked.
➤ Key release: Press decimal point key, then
key below END.
Loudness control of alarm tone
(only software IFPE)
The loudness of the audible alarm can be
adjusted in 9 steps.
➤ Increase with key below (+) resp. decrease
with key below (-).
➤ Exit selection: Press key below END.
Set Date/Clock
➤ Set date: Press key below DAT, enter date
and confirm with DAT.
➤ Set time: Press key below TIME, enter time
and confirm with TIME.
➤ Exit selection: Press key below END.
Special Functions
10